不惜一切保留活动度?多节段铰链不愈合、钢板断裂及硬膜内钢板移位在颈椎椎板成形术后的病例报告与文献综述
本文报告一例61岁男性颈椎椎板成形术后7年出现多节段铰链不愈合、钢板断裂及C7钢板碎片硬膜内移位,表现为严重头痛和左上、下肢电击样疼痛;作者据此提出预防铰链不愈合的建议,因微动可能促成钢板断裂。
要点
- 颈椎椎板成形术保留活动度,常见并发症包括C5神经根麻痹、轴性颈痛、铰链不愈合和过早闭合。
- 钢板固定较缝线等刚性更小的固定可能提供更好的术后稳定并降低椎板成形术后闭合风险。
- 本例出现多节段铰链不愈合、钢板断裂及硬膜内钢板碎片移位,作者提出预防铰链不愈合的建议。
核心结论
- 作者提出预防铰链不愈合的建议,因为由此产生的微动可能促成该患者观察到的钢板断裂。
研究设计与样本
- 病例报告及文献综述。
主要发现
- 61岁男性,7年前因颈椎病性脊髓病行C3-7椎板成形术,因严重头痛及左上、下肢电击样疼痛就诊。
- 影像学显示多块椎板成形钢板断裂及C7钢板碎片硬膜内移位。
临床怎么用
- 作者提出预防铰链不愈合的建议,因微动可能促成钢板断裂;原文未据此给出其他临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Cervical laminoplasty is a motion-preserving procedure that addresses spinal cord compression and avoids postlaminectomy kyphosis associated with cervical laminectomy. The most common complications include C5 nerve palsy, axial neck pain, hinge nonunion, and premature closure. Plating is a relatively newer method of laminoplasty fixation that may provide greater stabilization postoperatively and reduce the risk of laminoplasty closure compared with less rigid (e.g., suture) fixation techniques. Although prior studies have reported low rates of laminar/lateral mass screw back out, plate breakage and migration have not been previously described in the literature. The purpose of this paper is to present a case of multilevel hinge nonunion, plate breakage, and plate fragment migration. Although rare, plate failure may result in a dural tear and spinal cord injury/compression. CASE DESCRIPTION: In this case, a 61-year-old man with a history of cervical spondylotic myelopathy treated with C3-7 laminoplasty 7 years prior presented to our hospital with severe headaches and electrical-type pain through the left upper and lower extremities. Imaging studies revealed several broken laminoplasty plates and intradural migration of a fragment of the C7 plate. CONCLUSIONS: We provide recommendations for preventing hinge nonunion because resultant micromotion likely contributed to the plate breakages observed in this patient.
引用
- Rosinski A, Odeh K, Leasure J, et al. Motion Preservation at All Costs? Multilevel Hinge Nonunion, Plate Breakage, and Intradural Plate Migration After Cervical Laminoplasty: A Case Report and Literature Review. World Neurosurg. 2019. PMID: 31759152. doi:10.1016/j.wneu.2019.11.074
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。