该回顾性研究来自2家机构,纳入11例患者共12个颅内动脉瘤,在超说明书部位单独或联合支架/弹簧圈使用WEB装置。动脉瘤位于后交通动脉、眼动脉、胼周动脉、小脑后下动脉、椎基底动脉交界及海绵窦段颈内动脉。无手术相关死亡或症状性并发症,作者认为疗效需多中心前瞻性研究评估,但初步结果有前景。
要点
- 11例患者12个动脉瘤,4个(30.8%)为破裂动脉瘤,均在海绵窦段颈内动脉等超说明书部位使用WEB装置。
- 术后完全闭塞4例(33.3%),充分闭塞8例(66.6%),无手术相关死亡或症状性并发症。
- 作者认为该用法疗效需多中心前瞻性研究评估,但初步结果有前景。
核心结论
- WEB装置可成功部署于后交通动脉、眼动脉及胼周动脉等超说明书部位动脉瘤。
- 作者认为该用法的疗效需在多中心前瞻性研究中评估,但初步结果有前景。
- 作者指出准确描述动脉瘤大小/形态以及了解可能的辅助治疗对成功干预至关重要。
研究设计与样本
- 来自2家机构的回顾性研究,纳入11例颅内动脉瘤患者共12个动脉瘤,接受WEB装置单独或联合支架和/或弹簧圈在超说明书部位治疗。
主要发现
- 12个动脉瘤中4个(30.8%)为破裂动脉瘤。
- 超说明书部位包括后交通动脉3个、眼动脉眶上段3个、胼周动脉2个、小脑后下动脉2个、椎基底动脉交界1个、海绵窦段颈内动脉1个。
- 平均瘤顶-瘤颈比为1.91 ± 0.9,手术平均时长63.9 ± 29.6分钟。
- 股动脉入路占61.5%,桡动脉入路占48.5%;5例手术(41.7%)在监测麻醉下进行;每个动脉瘤平均尝试1.33次。
- WEB栓塞辅助弹簧圈2例(16.67%),辅助支架1例(8.3%)。
临床怎么用
- 作者认为准确描述动脉瘤大小/形态以及了解可能的辅助治疗对成功干预至关重要。
- 作者认为该用法的疗效需在多中心前瞻性研究中评估,但初步结果有前景。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: We present our initial experience with the off-label use of the Woven EndoBridge (WEB) device. METHODS: We performed a retrospective study from 2 institutions of patients with intracranial aneurysms who underwent treatment with the WEB device alone or in conjunction with stenting and/or coiling in an off-labeled location. RESULTS: Eleven patients with 12 aneurysms were included. Four (30.8%) aneurysms presented ruptured. Off-labeled locations included 3 aneurysms in the posterior communicating artery, 3 in the supraclinoid ophthalmic artery, 2 in the pericallosal artery, 2 in the posterior inferior cerebellar artery, 1 at the vertebrobasilar junction, and 1 in the cavernous internal carotid artery. Mean dome-to-neck ratio was 1.91 ± 0.9. Average duration of the procedure was 63.9 ± 29.6 minutes. Femoral route was used in 61.5%, and radial access in 48.5% of cases. Five procedures (41.7%) were performed under monitored anesthesia care. We had 1.33 attempts per aneurysm. WEB embolization was supplemented with coiling in 2 cases (16.67%) and stenting in 1 case (8.3%). Based on the O'Kelly-Marotta ("OKM") grading scale, we report a rate of 0% for grade 1, 33.3% for grade 2, and 33.3% for grade 3. Postprocedural occlusion was complete in 4 (33.3%) and adequate in 8 (66.6%) patients. No procedure-related deaths or symptomatic complications were reported. CONCLUSIONS: The WEB device was successfully deployed in off-labeled locations, including posterior communicating artery, ophthalmic artery, and pericallosal aneurysms. The efficacy of this usage must be evaluated in multicenter prospective studies, but our preliminary results are promising. Correct characterization of the aneurysm size/morphology, as well as awareness of possible adjunct treatments, are crucial for successful intervention.
引用
- Zanaty M, Roa JA, Tjoumakaris SI, et al. Off-Label Use of the WEB Device. World Neurosurg. 2019. PMID: 31759154. doi:10.1016/j.wneu.2019.11.076
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。