石墨涂层与远端裂隙导管梗阻的炎症机制
该研究回顾22例手术取出的腹膜分流导管和8例尸检病例,发现多数症状性梗阻与远端裂隙导管相关,早期组织反应为T淋巴细胞聚集和巨噬细胞吞噬石墨颗粒,继而纤维母细胞、间皮细胞和血管增生穿过裂隙导致管腔阻塞,约1年后组织栓子胶原化钙化。
要点
- 远端裂隙导管的石墨涂层可能诱发炎症反应并导致梗阻。
- 早期组织反应以T淋巴细胞聚集和巨噬细胞吞噬石墨颗粒为特征。
- 作者建议除非有对照试验证明安全有效,否则不应使用裂隙阀远端导管。
核心结论
- 石墨涂层可能使远端裂隙导管易于发生梗阻。
- 神经外科医生和制造商应考虑该分流设计潜在的负面后果。
- 作者同意既往建议:除非对照试验证明安全有效,否则不应使用裂隙阀远端导管进行脑室腹腔分流。
研究设计与样本
- 回顾性质量保证审查,详细调查22例手术取出的腹膜分流导管和8例尸检病例。
- 对导管和/或组织进行常规组织学和免疫组化评估,其中3例进行电子显微镜检查及元素分析。
主要发现
- 多数症状性梗阻与远端裂隙导管相关(17例裂隙、3例开放端、2例类型未知)。
- 尸检病例中,死亡归因于分流失败:2例裂隙导管梗阻、1例开放端导管梗阻、1例导管从腹膜腔撤出。
- 早期组织反应主要为T淋巴细胞聚集,巨噬细胞吞噬石墨颗粒,伴随纤维母细胞、间皮细胞和血管增生穿过裂隙阻塞管腔。
- 炎症约1年后消退,组织栓子胶原化并钙化。
临床怎么用
- 避免使用远端裂隙导管进行脑室腹腔分流,除非有对照试验证明其安全有效。
- 对使用裂隙导管的分流患者,应警惕早期梗阻可能。
局限性
- 回顾性研究,样本量小(22例导管和8例尸检)。
- 原文未报告对照试验数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Tissue reactions that contribute to obstruction of peritoneal catheters in ventriculoperitoneal shunt systems are not well characterized. Several recent rapid obstructions in children prompted a retrospective quality assurance review. METHODS: The authors conducted a detailed investigation of 22 surgically explanted peritoneal shunt catheters and 8 autopsy cases with documented distal shunt obstruction. Patients' medical histories were reviewed, and the catheters and/or tissues were subjected to conventional histological and immunohistochemical evaluations. In addition, 3 cases were subjected to electron microscopic examination including elemental analysis. RESULTS: The majority of symptomatic obstructions were associated with distal slit catheters (17 slit, 3 open-end, and 2 unknown type). Among the autopsy cases, deaths were attributed to shunt failure in 2 cases of slit catheter obstruction, 1 case of open-end catheter obstruction, and 1 case of catheter withdrawal from the peritoneal cavity. The early tissue response consisted of a predominantly T lymphocyte accumulation with phagocytosis of graphite particles by macrophages. This is associated with proliferation of fibroblasts, mesothelial cells, and blood vessels, which can grow through the slits and occlude the catheter lumen. As the inflammation subsides after approximately 1 year, the tissue plug becomes collagenized and calcified. CONCLUSIONS: This study, supported by experimental literature in other organ systems, indicates that graphite used to coat the slit openings of distal catheters from ventriculoperitoneal shunts likely predisposes to obstruction. Neurosurgeons and manufacturers should consider the potential negative consequences of this shunt design. The authors concur with previous recommendations that slit-valve distal catheters should not be used for ventriculoperitoneal shunting unless they can be proven safe and efficacious in a controlled trial.
引用
- Del Bigio MR, Sidhu RK, Kazina CJ, et al. Inflammation and obstruction of distal catheter slits in ventriculoperitoneal shunts: likely role of graphite. J Neurosurg. 2019. PMID: 31561214. doi:10.3171/2019.6.JNS191082
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。