超小阀门植入治疗新生儿脑积水
本研究回顾性分析2001年至2017年间单中心156例1岁以下接受脑室分流术的脑积水患儿,比较超小阀门与常规尺寸阀门的并发症率。两组2年内总体并发症率分别为37.5%和41.7%,1年分流存活率无差异。结论认为阀门尺寸不影响并发症率,但分流故障病因不同。
要点
- 1岁以下脑积水患儿中,超小阀门与常规阀门的2年并发症率无显著差异。
- 超小阀门组患儿平均年龄更小(2.1月 vs 3.1月,P=0.03)。
- 两组1年分流存活率无差异,但分流故障病因不同。
核心结论
- 超小阀门与常规阀门在1岁以下脑积水患儿中并发症率无显著差异,但分流故障病因不同。
研究设计与样本
- 单中心回顾性研究,纳入2001年1月至2017年1月接受脑室分流术的1岁以下患儿156例。
主要发现
- 2年内总体并发症率:超小阀门组37.5%,常规阀门组41.7%。
- 1年分流存活率两组无差异。
- 超小阀门组平均年龄2.1个月,常规阀门组3.1个月(P=0.03)。
临床怎么用
- 阀门尺寸选择可能不影响1岁以下脑积水患儿的总体并发症风险,但需注意故障病因差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Neonatal hydrocephalus remains a difficult condition to manage, due to high failure rates among all management strategies. Neurosurgeons commonly manage hydrocephalus with ventriculoperitoneal shunt (VPS) implantation, and valves of variable sizes and profiles are available for implantation. This study examines primary ventricular shunt valve implantation complication rates based on valve profiles in pediatric patients with hydrocephalus. METHODS: This study retrospectively reviews pediatric patients younger than 1 year of age who underwent ventricular shunt placement at a single institution from January 2001 to January 2017. Patients were classified by valve profile and categorized as either ultrasmall valves or regular-sized valves. Time until complication and type of complication were studied. RESULTS: A total of 156 patients met the inclusion criteria. Forty-eight (31%) patients received an ultrasmall shunt valve, while 108 patients received a regular valve. On average, patients undergoing ultrasmall valve placement were younger (2.1 months) than patients undergoing placement of regular valves (3.1 months) (P = 0.03). The overall complication rate within 2 years of VPS placement was 37.5% in patients with the ultrasmall valve and 41.7% in the regular valve population. There was no difference in 1-year shunt survival rate between the 2 cohorts. CONCLUSION: Our review did not find a significant difference in complication rates between ultrasmall and regular valves in patients under 1 year of age. However, the etiology of shunt malfunction did differ between the groups. This work further supports evidence suggesting a surgeon's preference for shunt hardware alone does not significantly impact outcome.
引用
- Reed SW, Cools MJ, Quinsey CS, et al. Neonatal Hydrocephalus Treatment with Ultrasmall Valve Implantation. World Neurosurg. 2019. PMID: 31526889. doi:10.1016/j.wneu.2019.09.043
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。