窄带增强激发提升常规神经外科显微镜的近红外术中分子成像
本研究在常规神经外科显微镜和近红外模块基础上,加用805nm激光增强吲哚菁绿(ICG)激发,体外对ICG连续稀释液定量,并对11例脑肿瘤(脑膜瘤、胶质母细胞瘤、转移瘤)患者行延迟24小时ICG成像。结果显示增强激发使体外平均信背比(SBR)提高29%,患者术中肿瘤SBR平均改善35%(p < 0.001)。
要点
- 加用805nm激光增强激发可提高常规显微镜的近红外荧光信背比。
- 体外ICG连续稀释液平均SBR提高29%。
- 11例脑肿瘤患者术中肿瘤SBR平均改善35%(p < 0.001)。
核心结论
- 在常规神经外科显微镜的近红外模块上加用聚焦激发源,可简单有效地改善脑肿瘤近红外可视化。
研究设计与样本
- 体外研究对ICG连续稀释液进行定量,并对11例脑膜瘤、胶质母细胞瘤和转移瘤患者行延迟24小时ICG输注的肿瘤切除及术中成像。
主要发现
- 体外增强激发使六种ICG连续稀释液的平均SBR提高29%。
- 11例患者中,增强激发使肿瘤SBR较增强前平均改善35%(p < 0.001)。
- 直接肿瘤可视化下肿瘤荧光增强明显。
关键数值与推荐
- 增强激发后体外平均SBR提高29%。
- 患者术中增强激发后SBR平均改善35%(p < 0.001)。
临床怎么用
- 常规神经外科显微镜仍是颅内手术肿瘤可视化的首选方法,但现有近红外信号检测平台欠佳,加用聚焦激发源可改善信号。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Intraoperative visualization of brain tumors with near-infrared (NIR)-fluorescent dyes is an emerging method for tumor margin approximation but are limited by existing fluorescence detection platforms. We previously showed that a dedicated NIR imaging platform outperformed a state-of-the-art neurosurgical microscope in fluorescence signal characteristics. This study examined whether conventional neurosurgical microscope NIR signal could be improved with the addition of a narrow wavelength excitation source. METHODS: Imaging was conducted with a broad-spectrum neurosurgical microscope and commercial near-infrared module. Addition of an 805-nm laser was used to "boost" NIR excitation of indocyanine green (ICG). In vitro quantification was performed on serial dilutions of ICG. Patients underwent tumor resection with delayed 24-h imaging of ICG infusion. NIR fluorescence of dura, cortex, or tumor was quantified from images prior to (pre-boost) and following added excitation with the laser (post-boost). Signal to background ratio (SBR) of pre- and post-boost was calculated as a readout of image enhancement. RESULTS: In vitro, excitation boost effected a 29% increase in mean SBR in six serial dilutions of ICG. Intraoperative boost was performed in 11 patients including meningioma, glioblastoma multiforme, and metastases. Increase in tumor fluorescence was pronounced under direct tumor visualization. Across all patients, boost excitation resulted in 35% mean improvement from pre-boost SBR (p < 0.001). CONCLUSION: Neurosurgical microscopes remain the preferred method of visualizing tumor during intracranial surgery. However, current modalities for NIR signal detection are suboptimal. We demonstrate that augmentation of a fluorescence microscope module with a focused excitation source is a simple mechanism of improving NIR tumor visualization. CLINICAL TRIAL REGISTRATION: NCT03262636.
引用
- Li C, Buch L, Cho S, et al. Near-infrared intraoperative molecular imaging with conventional neurosurgical microscope can be improved with narrow band "boost" excitation. Acta Neurochir (Wien). 2019. PMID: 31482242. doi:10.1007/s00701-019-04054-5