多中心开放标签对照试验:可充电脑深部电刺激脉冲发生器接受度、便利性与并发症
这项多中心、患者聚焦、非企业主导的研究纳入195例运动障碍患者(帕金森病、震颤或肌张力障碍),评估可充电神经刺激器的充电便利性、充电负担与并发症。总体便利性评为「容易」,平均每周充电122分钟,93.8%用户对充电有信心,手术翻修率4.1%,感染率2.1%。
要点
- 总体充电便利性评为「容易」,93.8%用户对充电装置有信心。
- 平均每周充电负担122分钟,与治疗时长呈正相关。
- 手术翻修率4.1%,感染率2.1%,充电失败发生率8.7%。
核心结论
- 运动障碍患者总体认为可充电神经刺激器充电容易,并发症率低且充电负担可接受。
研究设计与样本
- 多中心、开放标签、对照试验,四家神经外科中心向所有成年运动障碍且植入可充电神经刺激器的患者发放问卷。
- 共回收195例患者数据集(占发放问卷的66.1%),患者平均年龄61.3岁,装置平均植入时长40.3个月。
- 主要终点为充电过程便利性,采用序数量表从「非常困难」(1)到「非常容易」(5)评分。
主要发现
- 总体充电便利性评分为「容易」(4)。
- 平均每周充电负担为122分钟,且与治疗时长呈正相关。
- 93.8%的用户对充电装置有信心。
- 手术翻修率为4.1%,感染率为2.1%。
- 8.7%的患者发生充电失败,3.6%的患者因充电失败导致治疗中断。
关键数值与推荐
- 帕金森病患者的便利性评分显著差于肌张力障碍患者。
- 12.3%的病例由照护者代为充电。
- 从非可充电装置转换为可充电装置的患者,相比初次植入即为可充电装置的患者,认为充电显著更不便且充电负担更高。
- 年龄对任何终点均无显著影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Rechargeable neurostimulators for deep brain stimulation have been available since 2008, promising longer battery life and fewer replacement surgeries compared to non-rechargeable systems. Long-term data on how recharging affects movement disorder patients are sparse. This is the first multicenter, patient-focused, industry-independent study on rechargeable neurostimulators. METHODS: Four neurosurgical centers sent a questionnaire to all adult movement disorder patients with a rechargeable neurostimulator implanted at the time of the trial. The primary endpoint was the convenience of the recharging process rated on an ordinal scale from "very hard" (1) to "very easy" (5). Secondary endpoints were charge burden (time spent per week on recharging), user confidence, and complication rates. Endpoints were compared for several subgroups. RESULTS: Datasets of 195 movement disorder patients (66.1% of sent questionnaires) with Parkinson's disease (PD), tremor, or dystonia were returned and included in the analysis. Patients had a mean age of 61.3 years and the device was implanted for a mean of 40.3 months. The overall convenience of recharging was rated as "easy" (4). The mean charge burden was 122 min/wk and showed a positive correlation with duration of therapy; 93.8% of users felt confident recharging the device. The rate of surgical revisions was 4.1%, and the infection rate was 2.1%. Failed recharges occurred in 8.7% of patients, and 3.6% of patients experienced an interruption of therapy because of a failed recharge. Convenience ratings by PD patients were significantly worse than ratings by dystonia patients. Caregivers recharged the device for the patient in 12.3% of cases. Patients who switched from a non-rechargeable to a rechargeable neurostimulator found recharging to be significantly less convenient at a higher charge burden than did patients whose primary implant was rechargeable. Age did not have a significant impact on any endpoint. CONCLUSIONS: Overall, patients with movement disorders rated recharging as easy, with low complication rates and acceptable charge burden.
引用
- Jakobs M, Helmers AK, Synowitz M, et al. A multicenter, open-label, controlled trial on acceptance, convenience, and complications of rechargeable internal pulse generators for deep brain stimulation: the Multi Recharge Trial. J Neurosurg. 2019. PMID: 31419794. doi:10.3171/2019.5.JNS19360
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。