微电极记录对运动障碍疾病脑深部电刺激电极位置的影响
该研究纳入273个行微电极记录(MER)的脑深部电刺激(DBS)手术半球,比较初始解剖靶点、MER优化靶点与最终电极位置。52%半球存在影像与神经生理靶点不一致,MER使电极向优化靶点显著偏移(2.11 vs 1.09 mm,P<0.05),但实际偏移量平均较小。
要点
- 52%的DBS半球存在影像与神经生理靶点不一致。
- 不一致半球中MER引导电极显著向优化靶点偏移。
- 电极偏移方向与MER优化方向一致,前外侧方向除外。
核心结论
- 影像与神经生理靶点不一致见于52%的半球,MER使DBS电极向适当靶点偏移,但实际偏移值平均较小。
研究设计与样本
- 连续纳入使用术中CT的MER辅助DBS手术病例,分析273个半球的初始解剖靶点、MER优化靶点与电极坐标。
主要发现
- 273个半球中143个(52%)为不一致半球,130个(48%)为一致半球。
- 一致半球中DBS电极放置误差为0.88±0.07 mm(均值±标准误)。
- 不一致半球中初始靶点与电极平均误差为2.11±0.07 mm,显著大于MER优化靶点与电极距离(2.11 vs 1.09 mm,P<0.05)。
- 不一致半球中电极按MER优化方向偏移,但预期方向为前外侧时例外。
临床怎么用
- MER可识别影像靶点与神经生理靶点不一致的半球,并引导电极向优化靶点调整。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: During deep brain stimulation (DBS) surgery, microelectrode recording (MER) leads to target refinement from the initial plan in 30% to 47% of hemispheres; however, it is unclear whether the DBS lead ultimately resides within the MER-optimized target in relation to initial radiographic target coordinates in these hemispheres. This study aimed to determine the frequency of discordance between radiographic and neurophysiologic nucleus and whether target optimization with MER leads to a significant change in DBS lead location away from initial target. METHODS: Consecutive cases of DBS surgery with MER using intraoperative computed tomography were included. Coordinates of initial anatomic target (AT), MER-optimized target (MER-O) and DBS lead were obtained. Hemispheres were categorized as "discordant" (D) if there was a suboptimal neurophysiologic signal despite accurate targeting of AT. Hemispheres where the first MER pass was satisfactory were deemed "concordant" (C). Coordinates and radial distances between 1) AT/MER-O; 2) MER-O/DBS; and 3) AT/DBS were calculated and compared. RESULTS: Of the 273 hemispheres analyzed, 143 (52%) were D, and 130 (48%) were C. In C hemispheres, DBS lead placement error (mean ± standard error of the mean) was 0.88 ± 0.07 mm. In D hemispheres, MER resulted in significant migration of DBS lead (mean AT-DBS error 2.11 ± 0.07 mm), and this distance was significantly greater than the distance between MER-O and DBS (2.11 vs. 1.09 mm, P < 0.05). Directional assessment revealed that the DBS lead migrated in the intended direction as determined by MER-O in D hemispheres, except when the intended direction was anterolateral. CONCLUSIONS: Discordance between radiographic and neurophysiologic target was seen in 52% of hemispheres, and MER resulted in appropriate deviation of the DBS lead toward the appropriate target. The actual value of the deviation, when compared with DBS lead placement error in C hemispheres, was, on average, small.
引用
- Kochanski RB, Bus S, Brahimaj B, et al. The Impact of Microelectrode Recording on Lead Location in Deep Brain Stimulation for the Treatment of Movement Disorders. World Neurosurg. 2019. PMID: 31449992. doi:10.1016/j.wneu.2019.08.092