用于清醒开颅术中高分辨率皮质脑电记录的360°环形电极装置
本研究评估一种360°高密度环形皮质电极装置(circular grid)在清醒开颅手术中同步切除肿瘤并实时记录皮质脑电的价值。结论认为该装置安全可行,可能提高致痫灶检测准确性和功能预后。
要点
- 环形电极为清醒开颅术中同步肿瘤切除与皮质脑电监测提供直接通道。
- 配对分析中环形电极组切除程度更高(p=0.03),术后短暂运动恢复期约缩短6.6天(p≤0.05)。
- 两组术后癫痫发作无显著差异(p=0.80),条状电极组KPS评分下降更明显(p=0.007)。
核心结论
- 环形电极装置安全可行,可在组织切除的同时直接监测皮质电活动。
- 应用于不同脑病理可能提高术中致痫灶检测准确性和功能预后,并减少术后并发症。
研究设计与样本
- 环形电极组29例、条状电极组70例;其中50例进行配对分析(环形电极25例、条状电极25例)。
主要发现
- 配对分析显示环形电极组切除程度更高(p=0.03)。
- 环形电极组住院期间短暂运动恢复期约缩短6.6天(p≤0.05)。
- 条状电极组术后较术前KPS评分差值/下降更大(p=0.007)。
- 两组术后癫痫发作无显著差异(p=0.80)。
关键数值与推荐
- 幕上脑肿瘤患者中癫痫常见,约40%-70%的胶质瘤患者发生脑肿瘤相关癫痫(BTRE)。
- 致痫信号准确定位需要高密度空间组织的电极,并须在术中置于皮质表面。
临床怎么用
- 可在清醒开颅切除肿瘤时使用环形高密度电极同步记录皮质脑电,辅助致痫灶定位。
- 该装置可能有助于提高切除程度并缩短术后短暂运动恢复期,但需结合患者具体情况选择。
局限性
- 为回顾性研究,样本来自单中心,配对分析仅50例。
- 原文未报告随机化、盲法及长期随访结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Epilepsy is common among patients with supratentorial brain tumors; approximately 40%-70% of patients with glioma develop brain tumor-related epilepsy (BTRE). Intraoperative localization of the epileptogenic zone during surgical tumor resection (real-time data) may improve intervention techniques in patients with lesional epilepsy, including BTRE. Accurate localization of the epileptogenic signals requires electrodes with high-density spatial organization that must be placed on the cortical surface during surgery. The authors investigated a 360° high-density ring-shaped cortical electrode assembly device, called the "circular grid," that allows for simultaneous tumor resection and real-time electrophysiology data recording from the brain surface. METHODS: The authors collected data from 99 patients who underwent awake craniotomy from January 2008 to December 2018 (29 patients with the circular grid and 70 patients with strip electrodes), of whom 50 patients were matched-pair analyzed (25 patients with the circular grid and 25 patients with strip electrodes). Multiple variables were then retrospectively assessed to determine if utilization of this device provides more accurate real-time data and improves patient outcomes. RESULTS: Matched-pair analysis showed higher extent of resection (p = 0.03) and a shorter transient motor recovery period during the hospitalization course (by approximately 6.6 days, p ≤ 0.05) in the circular grid patients. Postoperative versus preoperative Karnofsky Performance Scale (KPS) score difference/drop was greater for the strip electrode patients (p = 0.007). No significant difference in postoperative seizures between the 2 groups was present (p = 0.80). CONCLUSIONS: The circular grid is a safe, feasible tool that grants direct access to the cortical surgical surface for tissue resection while simultaneously monitoring electrical activity. Application of the circular grid to different brain pathologies may improve intraoperative epileptogenic detection accuracy and functional outcomes, while decreasing postoperative complications.
引用
- ReFaey K, Chaichana KL, Feyissa AM, et al. A 360° electronic device for recording high-resolution intraoperative electrocorticography of the brain during awake craniotomy. J Neurosurg. 2019. PMID: 31277069. doi:10.3171/2019.4.JNS19261
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。