患者特异性内镜胶样囊肿切除手术模拟器的开发与评估
OBJECTIVE: Endoscopic resection of third-ventricle colloid cysts is technically challenging due to the limited dexterity and visualization provided by neuroendoscopic instruments. Extensive training and experience are required to master the learning curve.…
要点
- 患者特异性合成模拟器成功开发,用于内镜胶样囊肿切除训练。
核心结论
- 患者特异性合成手术模拟器成功开发,可用于内镜胶样囊肿切除训练。
- 模拟器的解剖、器械操作和程序内容被认为具有真实性。
- 模拟器可能有助于学习关键步骤、理解脑室解剖和掌握双手内镜技术。
研究设计与样本
- 基于8岁胶样囊肿伴脑积水患者的轴位平扫CT和T1加权钆增强MRI开发模拟器。
- 使用图像分割、计算机辅助设计、快速原型(3D打印)和硅胶模塑技术制作颅骨、脑、脑室和胶样囊肿模型。
- 采用19项反馈调查(5点Likert量表)评估解剖、器械操作、程序内容、感知真实性和信心与舒适度5个领域。
主要发现
- 解剖真实性平均评分4.34 ± 0.63(标准差)。
- 实际器械使用平均评分4.38 ± 0.58。
- 程序内容平均评分4.28 ± 0.77。
- 感知真实性平均评分4.08 ± 0.63。
关键数值与推荐
- 建议为经验丰富的受训者开发复杂病例场景(正常大小脑室、脉络丛粘连于囊肿壁、出血场景)。
- 建议纳入先进器械如侧切吸引装置。
临床怎么用
- 模拟器可作为学习内镜胶样囊肿切除关键步骤的教育工具。
- 有助于培养对脑室内解剖的详细理解和双手神经内镜技术的熟练度。
局限性
- 感知真实性评分略低于其他领域。
- 受训者建议开发复杂病例场景和纳入先进器械。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Endoscopic resection of third-ventricle colloid cysts is technically challenging due to the limited dexterity and visualization provided by neuroendoscopic instruments. Extensive training and experience are required to master the learning curve. To improve the education of neurosurgical trainees in this procedure, a synthetic surgical simulator was developed and its realism, procedural content, and utility as a training instrument were evaluated. METHODS: The simulator was developed based on the neuroimaging (axial noncontrast CT and T1-weighted gadolinium-enhanced MRI) of an 8-year-old patient with a colloid cyst and hydrocephalus. Image segmentation, computer-aided design, rapid prototyping (3D printing), and silicone molding techniques were used to produce models of the skull, brain, ventricles, and colloid cyst. The cyst was filled with a viscous fluid and secured to the roof of the third ventricle. The choroid plexus and intraventricular veins were also included. Twenty-four neurosurgical trainees performed a simulated colloid cyst resection using a 30° angled endoscope, neuroendoscopic instruments, and image guidance. Using a 19-item feedback survey (5-point Likert scales), participants evaluated the simulator across 5 domains: anatomy, instrument handling, procedural content, perceived realism, and confidence and comfort level. RESULTS: Participants found the simulator's anatomy to be highly realistic (mean 4.34 ± 0.63 [SD]) and appreciated the use of actual instruments (mean 4.38 ± 0.58). The procedural content was also rated highly (mean 4.28 ± 0.77); however, the perceived realism was rated slightly lower (mean 4.08 ± 0.63). Participants reported greater confidence in their ability to perform an endoscopic colloid cyst resection after using the simulator (mean 4.45 ± 0.68). Twenty-three participants (95.8%) indicated that they would use the simulator for additional training. Recommendations were made to develop complex case scenarios for experienced trainees (normal-sized ventricles, choroid plexus adherent to cyst wall, bleeding scenarios) and incorporate advanced instrumentation such as side-cutting aspiration devices. CONCLUSIONS: A patient-specific synthetic surgical simulator for training residents and fellows in endoscopic colloid cyst resection was successfully developed. The simulator's anatomy, instrument handling, and procedural content were found to be realistic. The simulator may serve as a valuable educational tool to learn the critical steps of endoscopic colloid cyst resection, develop a detailed understanding of intraventricular anatomy, and gain proficiency with bimanual neuroendoscopic techniques.
引用
- Bodani VP, Breimer GE, Haji FA, et al. Development and evaluation of a patient-specific surgical simulator for endoscopic colloid cyst resection. J Neurosurg. 2019. PMID: 31252392. doi:10.3171/2019.4.JNS183184
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。