利用电灼烟雾的差分迁移谱分析识别脑肿瘤
研究评估了差分迁移谱(DMS)分析电灼烟雾以术中识别脑肿瘤的可行性及初步验证。在28例神经外科手术患者的离体样本中,将组织切成694份小标本,进行7类分类。总体分类准确率为50%(随机水平14%),排除Tissue-Tek保存液中的样本后准确率提升至83%,低级别胶质瘤(II级)与对照的二分类准确率最高达94%。
要点
- DMS分析电灼烟雾可区分肿瘤与非肿瘤组织及不同肿瘤类型。
- 排除Tissue-Tek保存样本后分类准确率显著提高至83%。
- 低级别胶质瘤(II级)与对照的二分类准确率最高,达94%。
核心结论
- 不同脑肿瘤的电灼烟雾具有 distinct 的DMS谱,DMS分析仪连接特殊采样系统可在离体条件下区分肿瘤与非肿瘤组织及不同肿瘤类型。
研究设计与样本
- 离体实验研究,使用连接特殊烟雾采样系统的DMS分析28例神经外科手术患者的脑肿瘤及对照样本。
- 样本包括脑膜瘤(WHO I级)、毛细胞星形细胞瘤(I级)、其他低级别胶质瘤(II级)、胶质母细胞瘤(IV级)、CNS转移瘤以及出血性或创伤性脑组织作为对照,共切成694份小标本。
主要发现
- 7类分类的总体分类准确率(CA)为50%(随机水平14%)。
- 排除最初保存在Tissue-Tek保存液中的样本后,CA显著提升至83%。
- 使用更少类别时CA进一步提高,低级别胶质瘤(II级)与对照的二分类准确率最高,为94%。
临床怎么用
- 该技术有望为神经外科提供实时术中组织识别,但需进一步在体验证及优化采样与保存流程。
局限性
- 研究为离体实验,样本保存介质(Tissue-Tek)影响分类准确率,需排除后准确率才显著提升。
- 样本量有限(28例患者),分类类别较多时准确率仅50%,临床适用性尚需验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: There is a need for real-time, intraoperative tissue identification technology in neurosurgery. Several solutions are under development for that purpose, but their adaptability for standard clinical use has been hindered by high cost and impracticality issues. The authors tested and preliminarily validated a method for brain tumor identification that is based on the analysis of diathermy smoke using differential mobility spectrometry (DMS). METHODS: A DMS connected to a special smoke sampling system was used to discriminate brain tumors and control samples ex vivo in samples from 28 patients who had undergone neurosurgical operations. They included meningiomas (WHO grade I), pilocytic astrocytomas (grade I), other low-grade gliomas (grade II), glioblastomas (grade IV), CNS metastases, and hemorrhagic or traumatically damaged brain tissue as control samples. Original samples were cut into 694 smaller specimens in total. RESULTS: An overall classification accuracy (CA) of 50% (vs 14% by chance) was achieved in 7-class classification. The CA improved significantly (up to 83%) when the samples originally preserved in Tissue-Tek conservation medium were excluded from the analysis. The CA further improved when fewer classes were used. The highest binary classification accuracy, 94%, was obtained in low-grade glioma (grade II) versus control. CONCLUSIONS: The authors' results show that surgical smoke from various brain tumors has distinct DMS profiles and the DMS analyzer connected to a special sampling system can differentiate between tumorous and nontumorous tissue and also between different tumor types ex vivo.
引用
- Haapala I, Karjalainen M, Kontunen A, et al. Identifying brain tumors by differential mobility spectrometry analysis of diathermy smoke. J Neurosurg. 2019. PMID: 31200382. doi:10.3171/2019.3.JNS19274
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。