三腔螺栓多模态监测在重型急性脑损伤中的安全性
本研究基于一级创伤中心的前瞻性观察数据库,回顾性分析了501例接受三腔螺栓置入颅内压、脑温和脑氧三合一监测的重型急性脑损伤患者。共置入696个三腔螺栓,中位监测时长78.88小时,装置相关小挫伤9例(2.13%)、轴外血肿10例(2.36%),仅1例感染,结论认为三腔螺栓多模态监测安全性良好,并发症率与已发表的单腔螺栓及单一监测装置相当。
要点
- 三腔螺栓可同时监测颅内压、脑温和脑氧,并发症率与单腔螺栓及单一监测装置相当。
- 装置相关轴外血肿10例(2.36%),其中4例需治疗,多与置入技术而非装置本身相关。
- 因怀疑功能异常而更换的35个监测器中,19个(54.29%)复查功能正常。
核心结论
- 使用三腔螺栓置入颅内监测器进行多模态神经监测似乎是安全的。
- 并发症率与已发表的单腔螺栓和单一监测器的并发症率相似。
研究设计与样本
- 来自一级创伤中心的前瞻性观察数据库,回顾性识别501例患者,其中男300例、女201例,平均年龄58±39岁。
- 每例患者均接受三腔螺栓及3个监测器:颅内压、脑温和脑氧。
- 共置入696个三腔螺栓,中位监测时长78.88小时(四分位距33.0-133.2小时)。
- 22例(3.16%)患者置入双侧监测器。
主要发现
- 10个(1.43%)监测器因需行磁共振成像而更换,40个(5.74%)因需行额外颅脑手术而更换。
- 35个(5.02%)监测器因被认为功能异常而更换,其中19个(54.29%)随后发现功能正常。
- 随访CT与置入前CT比较,发现装置相关小挫伤9例(2.13%)和轴外血肿10例(2.36%)。
- 基于CT表现,血肿被认为与置入技术而非装置相关,其中4个血肿需要治疗。
- 22个(3.16%)装置置入位置错误(如探头位于梗死区或已有挫伤区),仅发现1例相关感染。
临床怎么用
- 在重型急性脑损伤患者中可考虑使用三腔螺栓进行颅内压、脑温和脑氧的多模态监测。
- 置入操作技术可能与装置相关轴外血肿有关,需关注置入技巧。
- 怀疑监测器功能异常时,应核实其功能后再决定是否更换。
局限性
- 本研究为单中心回顾性分析,来自一级创伤中心的前瞻性观察数据库。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Multimodality monitoring is used frequently to guide care of patients with severe acute brain injury. The aim of this study was to examine the safety and reliability of multimodality monitoring. METHODS: From a prospective observational database at a Level I trauma center, 501 patients, including 300 men and 201 women (mean age 58 ± 39 years) were identified retrospectively. Each patient received a triple-lumen bolt and 3 monitors: intracranial pressure, brain temperature, and brain oxygen. Intensive care unit and hospital records were examined to identify complications, reasons for device replacement, malfunction and infection. Head computed tomography (CT) scans performed before and after the monitors were inserted were examined for evidence of monitor-related adverse effects. RESULTS: A total of 696 triple-lumen bolts were placed. Median duration of monitoring was 78.88 hours (interquartile range, 33.0-133.2 hours). Bilateral monitors were inserted in 22 (3.16%) patients. Ten (1.43%) monitors were replaced to allow magnetic resonance imaging, and 40 (5.74%) monitors were replaced to facilitate additional cranial surgery. Of 35 (5.02%) monitors that were replaced because they were thought to not be functioning properly, 19 (54.29%) were subsequently found to be functioning normally. Follow-up CT scans were compared with CT scans obtained before insertion of monitors; 9 (2.13%) small contusions and 10 (2.36%) extra-axial hematomas associated with the devices were identified. Based on the CT findings, the hematomas were thought to be associated with the insertion technique rather than the device; 4 hematomas required treatment. Twenty-two (3.16%) devices were incorrectly placed (e.g., the probe was in an infarct or an already existing contusion). Only 1 associated infection was identified. CONCLUSIONS: Placement of intracranial monitors for multimodality neuromonitoring using a triple-lumen bolt appears to be safe. The complication rate is similar to published complication rates for single-lumen bolts and single monitors.
引用
- Bailey RL, Quattrone F, Curtin C, et al. The Safety of Multimodality Monitoring Using a Triple-Lumen Bolt in Severe Acute Brain Injury. World Neurosurg. 2019. PMID: 31195129. doi:10.1016/j.wneu.2019.05.195
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。