文献速递小儿神经外科与脑积水功能神经外科

Wishbone 颅中线定位装置在脑室腹腔分流术中的应用

The Wishbone: A Cranial Midline Localizing Device.
PubMed 文献 · World Neurosurg · 2019-05-14

术后 CT 显示装置定位点与解剖中线平均相距 2.9 mm(95%CI 1.6-4.1 mm),全部病例导管进入同侧侧脑室。

要点

  • Wishbone 定位耗时少于 3 分钟,操作简便、机械稳定。
  • 装置定位点与 CT 解剖中线平均偏差 2.9 mm(95%CI 1.6-4.1 mm)。

核心结论

  • Wishbone 是一种简单、高效、低成本的颅后中线准确定位系统,但尚需更大样本与无框架立体定向方法做确证性比较。

主要发现

  • Wishbone 定位过程用时少于 3 分钟,装置机械稳定且术者易于使用。
  • Wishbone 定位的头皮中线点与 CT 定义解剖中线平均距离 2.9 mm(95%CI 1.6-4.1 mm)。
  • 全部病例脑室导管均进入同侧侧脑室,导管尖端位于同侧额角 34 例、对侧额角 5 例。
  • 1 例导管尖端进入脑实质,原因为钻孔头尾向定位误差,与 Wishbone 无关。

关键数值与推荐

  • 定位精度:装置定位点与 CT 解剖中线平均距离 2.9 mm(95%CI 1.6-4.1 mm)。
  • 操作时间:定位过程少于 3 分钟。

临床怎么用

  • 在后颅中线定位(如 VPS 钻孔定位)中可用 Wishbone 快速辅助定位,但钻孔头尾向位置仍需术者自行把握。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The Wishbone device is designed to enable surgeons to quickly and accurately localize the cranial midline. It is intended to be of particular use when localizing the burr hole site during posterior ventriculoperitoneal shunt (VPS) surgery. METHODS: The Wishbone is a simple mechanical device with 2 adjustable caliper arms that reversibly attach to a patient's left and right external auditory canals. The Wishbone's laser localizer illuminates the midline scalp. The Wishbone was used to localize the posterior midline in a pilot series of patients undergoing VPS surgery. Midline localization and ventricular catheter placement accuracy were determined using findings from postoperative computed tomography scans. RESULTS: The Wishbone is a mechanically robust device and proved easy for surgeons to use. Forty patients underwent VPS surgery using the Wishbone to localize the posterior midline. The localization procedure took less than 3 minutes. The average distance separating the Wishbone-localized midline scalp location and the computed tomography-defined anatomical midline was 2.9 mm (95% confidence interval 1.6-4.1 mm). In all cases, the ventricular catheter entered the ipsilateral lateral ventricle. The catheter tips were placed in the ipsilateral (n = 34) or contralateral (n = 5) frontal horn in all but 1 patient. In 1 patient, the catheter tip entered the parenchyma due to a burr hole localization error in the rostrocaudal dimension, unrelated to the Wishbone. CONCLUSIONS: We describe a simple, efficient, and cost-effective system for accurately localizing the posterior cranial midline. A larger patient series is required to definitively compare its clinical utility relative to frameless stereotaxis-based midline localization methods.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Zanaty M, Banu M, Flouty O, et al. The Wishbone: A Cranial Midline Localizing Device. World Neurosurg. 2019. PMID: 31100521. doi:10.1016/j.wneu.2019.05.046

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。