文献速递功能神经外科

MRI引导聚焦超声治疗特发性震颤:低焦点温度重复声照射的累积热剂量与病灶相关性

Accumulated thermal dose in MRI-guided focused ultrasound for essential tremor: repeated sonications with low focal temperatures.
PubMed 文献 · J Neurosurg · 2019-05-10

结果显示累积热剂量(ATD)可预测术后MRI病灶大小,且低温和高温队列在3个月和1年时震颤评分均较基线改善。

要点

  • 50°C-54°C重复声照射可累积足够热剂量形成病灶并抑制震颤至术后1年。
  • T2w MRI病灶(5.0±1.4 mm)平均比T1w(3.9±1.4 mm)大28%。
  • 与T2w和T1w病灶相关性最佳的ATD阈值分别为100和200 CEM43。

核心结论

  • 在MRgFUS丘脑毁损术中,50°C-54°C重复声照射可累积足够热剂量形成病灶并产生临床相关震颤抑制至术后1年。
  • 累积热剂量(ATD)可用于预测MRI所测消融区大小。

研究设计与样本

  • 其中19例因颅骨条件不佳和/或颅骨加热相关疼痛而无法将焦点温度升至54°C以上,接受50°C-54°C重复声照射(5.1±1.5次)。

主要发现

  • 术后1天T2w MRI病灶(5.0±1.4 mm)平均比T1w MRI病灶(3.9±1.4 mm)大28%。
  • 与T2w病灶相关性最佳的ATD阈值为100 CEM43(回归斜率=0.97,R²=0.66)。
  • 与T1w病灶相关性最佳的ATD阈值为200 CEM43(回归斜率=0.98,R²=0.89)。
  • 低温(50°C-54°C)和高温(≥55°C)队列的CRST优势震颤子评分在术后3个月和1年均与基线有统计学差异。
  • 温度队列因素对优势震颤评分无显著固定效应。

关键数值与推荐

  • ATD阈值100 CEM43与T2w病灶相关性最佳(回归斜率=0.97,R²=0.66)。
  • ATD阈值200 CEM43与T1w病灶相关性最佳(回归斜率=0.98,R²=0.89)。

临床怎么用

  • 对于因颅骨条件或疼痛无法将焦点温度升至55°C以上的患者,可采用50°C-54°C重复声照射累积热剂量形成病灶。
  • 可利用累积热剂量预测术后MRI消融区大小,指导MRgFUS脑部手术规划。

局限性

  • 研究为单中心、特定患者亚组(19例低温重复声照射),样本量有限。
  • 热剂量累积前需对MR测温数据进行手动校正以补偿偏共振所致空间移位伪影。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The object of this study was to correlate lesion size with accumulated thermal dose (ATD) in transcranial MRI-guided focused ultrasound (MRgFUS) treatments of essential tremor with focal temperatures limited to 50°C-54°C. METHODS: Seventy-five patients with medically refractory essential tremor underwent MRgFUS thalamotomy at the authors' institution. Intraoperative MR thermometry was performed to measure the induced temperature and thermal dose distributions (proton resonance frequency shift coefficient = -0.00909 ppm/°C). In 19 patients, it was not possible to raise the focal temperature above 54°C because of unfavorable skull characteristics and/or the pain associated with cranial heating. In this patient subset, sonications with focal temperatures between 50°C and 54°C were repeated (5.1 ± 1.5, mean ± standard deviation) to accumulate a sufficient thermal dose for lesion formation. The ATD profile sizes (17, 40, 100, 200, and 240 cumulative equivalent minutes at 43°C [CEM43]) calculated by combining axial MR thermometry data from individual sonications were correlated with the corresponding lesion sizes measured on axial T1-weighted (T1w) and T2-weighted (T2w) MR images acquired 1 day posttreatment. Manual corrections were applied to the MR thermometry data prior to thermal dose accumulation to compensate for off-resonance-induced spatial-shifting artifacts. RESULTS: Mean lesion sizes measured on T2w MRI (5.0 ± 1.4 mm) were, on average, 28% larger than those measured on T1w MRI (3.9 ± 1.4 mm). The ATD thresholds found to provide the best correlation with lesion sizes measured on T2w and T1w MRI were 100 CEM43 (regression slope = 0.97, R2 = 0.66) and 200 CEM43 (regression slope = 0.98, R2 = 0.89), respectively, consistent with data from a previous study of MRgFUS thalamotomy via repeated sonications at higher focal temperatures (≥ 55°C). Two-way linear mixed-effects analysis revealed that dominant tremor subscores on the Fahn-Tolosa-Marin Clinical Rating Scale for Tremor (CRST) were statistically different from baseline at 3 months and 1 year posttreatment in both low-temperature (50°C-54°C) and high-temperature (≥ 55°C) patient cohorts. No significant fixed effect on the dominant tremor scores was found for the temperature cohort factor. CONCLUSIONS: In transcranial MRgFUS thalamotomy for essential tremor, repeated sonications with focal temperatures between 50°C and 54°C can accumulate a sufficient thermal dose to generate lesions for clinically relevant tremor suppression up to 1 year posttreatment, and the ATD can be used to predict the size of the resulting ablation zones measured on MRI. These data will serve to guide future clinical MRgFUS brain procedures, particularly those in which focal temperatures are limited to below 55°C.

来源:J Neurosurg,查看原文。版权归原作者所有。

引用

  1. Jones RM, Kamps S, Huang Y, et al. Accumulated thermal dose in MRI-guided focused ultrasound for essential tremor: repeated sonications with low focal temperatures. J Neurosurg. 2019. PMID: 31075781. doi:10.3171/2019.2.JNS182995

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。