PulseRider辅助弹簧圈栓塞治疗颅内分叉部动脉瘤:24个月随访单中心病例系列
单中心单臂注册研究前瞻性纳入8例未破裂颅内分叉部动脉瘤患者,使用PulseRider装置辅助弹簧圈栓塞,平均年龄66岁,动脉瘤直径4.6-13.6 mm,穹隆/颈比1.4-2.2。所有病例装置成功释放,6个月随访完全闭塞6/8(75%),24个月随访无动脉瘤复发或临床并发症,提示该装置治疗宽颈分叉部动脉瘤有效。
要点
- PulseRider辅助栓塞在8例分叉部动脉瘤中均成功释放,无操作相关并发症。
- 6个月随访完全闭塞率75%(6/8),近完全闭塞12.5%(1/8),残留12.5%(1/8)。
- 24个月随访无动脉瘤复发或临床并发症,mRS无变化。
核心结论
- 24个月随访显示PulseRider辅助弹簧圈栓塞治疗颅内宽颈分叉部动脉瘤具有良好的有效性。
研究设计与样本
- 单中心、单臂注册研究,经机构审查委员会批准,前瞻性纳入分叉部动脉瘤患者。
- 共8例患者,平均年龄66岁,动脉瘤包括2例前交通、2例颈动脉末端、4例基底动脉尖。
- 术后即刻及6个月行血管造影,12和24个月行MRI和MRA随访,回顾性分析影像。
- 术前、出院时及6、12、24个月随访获取改良Rankin评分。
主要发现
- 动脉瘤直径范围4.6-13.6 mm(平均7.4 mm),穹隆/颈比范围1.4-2.2(平均1.6)。
- 所有病例PulseRider均成功释放。
- 6个月随访完全闭塞6/8(75%),近完全闭塞1/8(12.5%),残留动脉瘤1/8(12.5%)。
- 24个月随访MRA未见变化或复发,临床无并发症。
关键数值与推荐
- 6个月随访完全闭塞率75%(6/8),近完全闭塞率12.5%(1/8),残留率12.5%(1/8)。
- 24个月随访无复发或临床并发症。
临床怎么用
- 对于宽颈分叉部动脉瘤,PulseRider辅助弹簧圈栓塞可作为治疗选择之一。
- 术后需规律影像随访(6个月造影,12和24个月MRA)评估闭塞情况。
局限性
- 单中心、单臂、样本量小(8例),缺乏对照组。
- 影像随访中血管造影仅至6个月,12和24个月为MRA,可能影响闭塞评估精度。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Although endovascular coiling of unruptured aneurysms is widely accepted, the endovascular treatment of wide-neck bifurcation aneurysms remains one of the most challenging morphologies. Our purpose was to describe our experience with 24-month follow-up for the treatment of unruptured intracranial bifurcation aneurysms using the PulseRider (Cerenovus, New Brunswick, NJ). METHODS: This study is a single-center, single-arm registry performed under institutional review board control to evaluate efficacy and safety of the PulseRider. Patients with bifurcation aneurysms were identified and enrolled prospectively. Angiography immediately after treatment and at 6 months, and magnetic resonance imaging and magnetic resonance angiography at 12- and 24-month follow-up were retrospectively analyzed. A modified Rankin score was obtained prior to procedure, at discharge, and at 6-, 12- and 24-month follow-up visits. RESULTS: Eight patients with a mean age of 66 years were treated with the PulseRider. All patients had bifurcation aneurysms (2 anterior communicating, 2 carotid terminus, and 4 basilar apex). The aneurysm diameters ranged from 4.6 to 13.6 mm (mean 7.4 mm) with dome/neck ratio ranging from 1.4 to 2.2 (mean 1.6). In all cases, the PulseRider was successfully deployed. Complete occlusion was demonstrated at 6-month follow-up on 6 of 8 (75%), near complete occlusion in 1 of 8 (12.5%), and residual aneurysm in 1 of 8 (12.5%) patients. There was no change or recurrence on magnetic resonance angiography, nor clinical complication after the procedure through 24-month follow-up. CONCLUSIONS: Our experience with 24-month follow-up demonstrated favorable efficacy in the treatment of intracranial wide-neck bifurcation aneurysms using the PulseRider.
引用
- Sakai N, Imamura H, Arimura K, et al. PulseRider-Assisted Coil Embolization for Treatment of Intracranial Bifurcation Aneurysms: A Single-Center Case Series with 24-Month Follow-up. World Neurosurg. 2019. PMID: 31042599. doi:10.1016/j.wneu.2019.04.177
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。