文献速递功能神经外科

入路角度影响机器人立体脑电图电极置入精度

Approach Angle Affects Accuracy in Robotic Stereoelectroencephalography Lead Placement.
PubMed 文献 · World Neurosurg · 2019-04-25

结果显示规划角度>30°的斜行轨迹径向误差显著高于<30°的正交轨迹(2.05 mm vs. 1.45 mm,P < 0.001)。提示规划轨迹时应避免大偏斜角度,若不可避免需留出更大安全边际。

要点

  • 规划角度>30°的轨迹径向误差显著高于<30°轨迹(2.05 mm vs. 1.45 mm,P < 0.001)。
  • ROSA机器人可提高立体定向手术速度与效率。
  • 规划时应避免高偏斜角度,不可避免时需确保更大安全边际。

核心结论

  • 规划轨迹与颅骨的夹角影响ROSA机器人立体脑电图电极置入的最终精度,高偏斜角度轨迹径向误差更大。

主要发现

  • 规划角度>30°的斜行轨迹径向误差为2.05 mm,显著高于规划角度<30°的正交轨迹的1.45 mm(P < 0.001)。

临床怎么用

  • 规划立体脑电图电极轨迹时应尽量避免高偏斜角度,若不可避免需留出更大安全边际。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: One third of patients with epilepsy fail to gain optimal control using antiepileptic drugs. New advances in epilepsy surgery have reshaped some diagnostic and therapeutic modalities into less invasive approaches. To understand the cortical epileptogenic networks, stereoelectroencephalography uses depth electrodes as a tool for invasive intracranial monitoring. These electrodes are now being placed using a robot instead of the traditional Talairach stereotactic grid and frame with comparable accuracy and the advantage of shortened surgical time. Whereas accuracy is important for correct cortical sampling and surgical morbidity, the factors that play a role in the deviation of the electrode's trajectory are not yet understood. The aim of this study was to determine if the angle the planned trajectory makes with the skull impacts the final accuracy of placement. METHODS: Twenty-five patients with 319 stereotactic trajectories were examined. Skew angle and accuracy measurements were computed based on preoperative and postoperative computed tomography scans fused with trajectories planned using ROSA robot software. RESULTS: Trajectories with a planned angle of >30° (oblique group) had significantly higher radial error (2.05 mm vs. 1.45 mm, P < 0.001) than trajectories with planned angles <30° (orthogonal group). CONCLUSIONS: The ROSA robotic system is a powerful tool that increases the speed and efficiency of stereotactic neurosurgery. Care should be taken when planning trajectories to avoid high skew angles. If a high skew angle is unavoidable, care should be taken to ensure a larger margin of safety.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Iordanou JC, Camara D, Ghatan S, et al. Approach Angle Affects Accuracy in Robotic Stereoelectroencephalography Lead Placement. World Neurosurg. 2019. PMID: 31028981. doi:10.1016/j.wneu.2019.04.143

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。