外部ICP传感器最佳零参考点研究

Best zero level for external ICP transducer.
PubMed 文献 · Acta Neurochir (Wien) · 2019-03-08

结果显示耳道和眉间在仰卧或侧位时代表脑中心,但不同头位间有12 mmHg变异。使用头顶最高点作为零参考会低估ICP,校正后可获得最准确读数。

要点

  • 耳道或眉间零参考在严格仰卧或侧位时代表脑中心。
  • 45°抬头和直立位分别高估脑中心ICP 4.8±0.8和5.6±0.5 mmHg。
  • 45°侧位低估脑中心ICP 6.3±1.0 mmHg。

核心结论

  • 耳道和眉间在严格仰卧或侧位时代表脑中心,但不同头位间有12 mmHg变异。
  • 室间孔在抬头时沿OM线,但转头时不沿。
  • 头顶最高点零参考低估ICP,校正后提供最准确读数。

研究设计与样本

  • 选取因头痛等原因转诊神经影像且3D MRI正常的患者。
  • 识别Monro、BC、O、EAM和OM线并投影至正中矢状或轴位图像。
  • 每例扫描评估仰卧、45°抬头、直立和45°侧位。
  • 测量皮肤至脑表面、BC和Monro的距离,以均值±SD和/或范围(毫米)表示。

主要发现

  • 45°抬头高估BC-ICP 4.8±0.8 mmHg,直立高估5.6±0.5 mmHg。
  • 45°侧位低估ICP-BC 6.3±1.0 mmHg。
  • Monro位于OM线中点吻侧45±5 mm,BC下24(18-31)mm、前13(8-17)mm。
  • 校正5.9或6.3 mmHg后,BC-ICP偏差≤1.8 mmHg,Monro-ICP≤0.9 mmHg。

关键数值与推荐

  • EAM或眉间零参考在严格仰卧或侧位时理想对应BC。
  • OM线在抬头时跟随Monro,但转头时不跟随。
  • 头顶最高点零参考低估脑表面、Monro和BC的ICP。
  • 校正后偏差:BC-ICP≤1.8 mmHg,Monro-ICP≤0.9 mmHg。

临床怎么用

  • 严格仰卧或侧位时可用EAM或眉间作为零参考对应BC。
  • 抬头或直立时需注意高估BC-ICP,侧位时低估。
  • 使用头顶最高点零参考时需校正低估,校正后读数最准确。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Continuous monitoring of intracranial pressure (ICP) was introduced in the 1950s. For correct ICP recordings, the zero-reference point for the external pressure gauge must be placed next to a head anatomical structure. We evaluated different anatomical points as zero reference for the ICP device at different head positions and their relation to brain centre (BC), foramen of Monro (Monro), and brain surface. METHODS: Patients referred for neuroimaging due to e.g. headache all having normal 3D MRI scans were selected. Monro, BC, Orbit(O), external auditory meatus (EAM), and orbito-meatal (OM) line were identified and projected to mid-sagittal, or axial images. Each scan was evaluated like lying supine, 45° head elevations, upright, and 45° lateral position. Distances from skin to brain surface, BC, and Monro were measured. All values are presented as mean ± SD and/or range in millimetre. For conversion to mmHg, millimetre was multiplied by 0.074. RESULTS: Twenty MRI scans were examined. A zero reference at EAM or glabella was ideal at BC when head was strict supine or in the lateral position. At 45° head elevation, an overestimation of the BC-ICP by 4.8 ± 0.8 and in upright 5.6 ± 0.5 mmHg was found, and 45° lateral underestimated ICP-BC by 6.3 ± 1.0 mmHg. Monro was situated 45 ± 5 mm rostral to the mid-OM line and 24 (18-31) mm inferior and 13 (8-17) mm in front of BC. A zero-reference point aligned with the highest point of the head underestimated BC-ICP and Monro-ICP. If the ICP reading was added 5.9 or 6.3 mmHg, respectively, a deviation from BC-ICP was ≤ 1.8 mmHg and Monro-ICP was ≤ 0.9 mmHg in all head positions. CONCLUSIONS: EAM and glabella are defined anatomical structures representing BC when strict supine or lateral but with 12 mmHg variation with different head positions used in clinical practice. The OM line follows Monro at head elevation, but not when the head is turned. When the highest external point on the head is used, ICP values at brain surface as well as Monro and BC are underestimated. This underestimation is fairly constant and, when corrected for, provides the most exact ICP reading.