清醒手术术中检查监测仪疗效的定量评估
本研究通过语言功能测试模拟器,比较使用与不使用清醒手术术中检查监测仪(IEMAS)时神经外科医生的测试总时长和医患信息交换次数。结果显示,使用IEMAS时测试总时长显著缩短(116.1 ± 23.1秒 vs. 147.8 ± 48.7秒,P < 0.02),信息交换次数显著减少(0.2 ± 0.6次 vs. 16.1 ± 15.6次,P < 0.02),表明IEMAS具有临床实用性。
要点
- 使用IEMAS时语言功能测试总时长显著短于不使用IEMAS时(116.1 ± 23.1秒 vs. 147.8 ± 48.7秒,P < 0.02)。
- 使用IEMAS时外科医生与检查者之间的信息交换次数显著少于不使用IEMAS时(0.2 ± 0.6次 vs. 16.1 ± 15.6次,P < 0.02)。
- IEMAS通过整合所有设备信息,可高效进行语言功能测试,具有临床实用性。
核心结论
- 使用IEMAS时测试总时长缩短,信息交换次数减少,证明IEMAS具有实用性。
研究设计与样本
- 研究采用语言功能测试模拟器,以消除症状等不确定因素,神经外科医生参与测试,比较使用与不使用IEMAS的情况。
主要发现
- 使用IEMAS时测试总时长显著短于不使用IEMAS时(116.1 ± 23.1秒 vs. 147.8 ± 48.7秒;P < 0.02)。
- 使用IEMAS时外科医生与检查者之间的信息交换次数显著少于不使用IEMAS时(0.2 ± 0.6次 vs. 16.1 ± 15.6次;P < 0.02)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: When brain tumors are located near the language area, a test to assess language function is required. During the test, it is practical to display combined information obtained from all the equipment so that the surgeon can confirm the patient's response to the tasks. We developed the intraoperative examination monitor for awake surgery (IEMAS) mainly to combine all information so that the language function test could be performed efficiently. The IEMAS has proved to be useful in clinical settings; however, no quantitative evaluation has been performed. This study aimed to demonstrate the clinical usefulness of the IEMAS through comparison of cases with and without IEMAS use in language function test simulation. METHODS: The language function test simulator was created to eliminate any uncertain factors, such as symptoms, which vary among patients. Neurosurgeons participated in the test, and the usefulness of the IEMAS was investigated. We analyzed test duration and number of information exchanges between surgeon and examiner. RESULTS: Total test duration with IEMAS use was significantly shorter than without IEMAS use (116.1 ± 23.1 seconds vs. 147.8 ± 48.7 seconds; P < 0.02). The number of information exchanges between surgeon and examiner was significantly lower with IEMAS use than without IEMAS use (0.2 ± 0.6 times vs. 16.1 ± 15.6 times; P < 0.02). CONCLUSIONS: We compared cases with and without IEMAS use. Total test duration decreased with IEMAS use, and number of information exchanges was reduced, thus demonstrating the usefulness of the IEMAS.
引用
- Fukutomi Y, Yoshimitsu K, Tamura M, et al. Quantitative Evaluation of Efficacy of Intraoperative Examination Monitor for Awake Surgery. World Neurosurg. 2019. PMID: 30825617. doi:10.1016/j.wneu.2019.02.069
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。