新型无框架颅骨安装球关节导向阵列在图像引导神经外科中的应用
该研究在9只非人灵长类动物中进行了15次MRI引导的套管插入,靶向壳核,以验证新型无框架颅骨安装球关节导向阵列(BJGA)的准确性。结果显示套管尖端平均偏差为1.18±0.60 mm,观察者内和观察者间相关系数均>0.85,表明BJGA可可靠靶向皮层下脑结构,准确性可与当前无框架立体定向系统相媲美。
要点
- BJGA在非人灵长类动物中靶向壳核的15次MRI引导套管插入,平均尖端偏差1.18±0.60 mm。
- 拓扑分析显示上、后、右方向偏差,观察者内和观察者间相关系数均>0.85。
- BJGA尺寸和多功能性结合简化工作流程,可能适用于多种颅内神经外科手术。
核心结论
- 数据表明BJGA可通过MRI引导可靠靶向皮层下脑结构,准确性可与当前无框架立体定向系统相媲美。
研究设计与样本
- 在9只非人灵长类动物(NHP)上进行15次MRI引导的套管插入,每次靶向壳核。
主要发现
- 由2名独立评审者测量的平均套管尖端偏差为1.18±0.60 mm(均值±标准差)。
- 拓扑分析显示上、后、右方向偏差,观察者内和观察者间相关系数均>0.85,表明观察者内和观察者间评估有效可靠。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Successful convection-enhanced delivery of therapeutic agents to subcortical brain structures requires accurate cannula placement. Stereotactic guiding devices have been developed to accurately target brain nuclei. However, technologies remain limited by a lack of MRI compatibility, or by devices' size, making them suboptimal for direct gene delivery to brain parenchyma. The goal of this study was to validate the accuracy of a novel frameless skull-mounted ball-joint guide array (BJGA) in targeting the nonhuman primate (NHP) brain. METHODS: Fifteen MRI-guided cannula insertions were performed on 9 NHPs, each targeting the putamen. Optimal trajectories were planned on a standard MRI console using 3D multiplanar baseline images. After cannula insertion, the intended trajectory was compared to the final trajectory to assess deviation (euclidean error) of the cannula tip. RESULTS: The average cannula tip deviation was 1.18 ± 0.60 mm (mean ± SD) as measured by 2 independent reviewers. Topological analysis showed a superior, posterior, and rightward directional bias, and the intra- and interclass correlation coefficients were > 0.85, indicating valid and reliable intra- and interobserver evaluation. CONCLUSIONS: The data demonstrate that the BJGA can be used to reliably target subcortical brain structures by using MRI guidance, with accuracy comparable to current frameless stereotactic systems. The size and versatility of the BJGA, combined with a streamlined workflow, allows for its potential applicability to a variety of intracranial neurosurgical procedures, and for greater flexibility in executing MRI-guided experiments within the NHP brain.
引用
- Sudhakar V, Mahmoodi A, Bringas JR, et al. Development of a novel frameless skull-mounted ball-joint guide array for use in image-guided neurosurgery. J Neurosurg. 2019. PMID: 30771782. doi:10.3171/2018.10.JNS182169