深部脑刺激患者的3-T MRI:线圈与脉冲序列的安全性评估
本研究通过体模模拟DBS设备(电极3387与IPG Activa PC 37601)在3-T MRI下的温度变化、IPG电压、SAR与B1+rms,并对41例全内置且激活的DBS患者行3-T MRI(T1与fMRI)。体模中3D SPGR、EPI、DTI与SWI序列下电极温升小于2°C,T2加权序列可能因SAR较高导致更高升温;41例患者无不良事件。作者认为在选定脉冲序列下3-T MRI对DBS患者似乎安全,常规T2加权序列应避免。
要点
- 体模中3D SPGR、EPI、DTI与SWI序列下DBS电极温升小于2°C,T2加权序列可能升温更高。
- 41例全内置激活DBS患者行3-T MRI(T1与fMRI)无不良事件。
- 作者认为常规T2加权序列应避免用于DBS患者。
核心结论
- 在本研究的实验条件下,使用选定脉冲序列对DBS患者行3-T MRI扫描似乎是安全的。
- 一般而言,DBS患者应避免使用常规方案的T2加权序列。
- 补充的3-T MRI体模安全数据提示,限制性低于本研究患者所用条件的成像条件(如使用发射体部多阵列接收线圈)也可能安全。
- 鉴于植入DBS神经调控设备与MRI系统之间的相互作用,这些发现仅适用于本研究的实验条件。
研究设计与样本
- 构建模拟DBS神经调控设备(DBS电极型号3387与IPG Activa PC型号37601)的体模并进行3-T MRI扫描。
- 体模研究记录最腹侧DBS电极触点温度变化、IPG电压、SAR与B1+rms,使用发射体部多阵列接收与正交发射-接收头线圈,涵盖从「最差」到「最常见」的多种DBS配置。
- 对41例全内置且激活的DBS患者使用正交发射-接收头线圈行3-T MRI(T1与fMRI),评估MR图像、神经系统检查与IPG阻抗稳定性。
主要发现
- 体模研究中,两种线圈在3D SPGR、EPI、DTI与SWI序列下DBS电极温升均小于2°C。
- 具有强烈射频脉冲的序列如T2加权序列可能因SAR较高而导致更高升温。
- IPG未关机并保持恒定刺激频率,其平均电压输出未改变。
- 41例DBS患者行3-T MRI无不良事件。
关键数值与推荐
- 体模中DBS电极温升小于2°C(3D SPGR、EPI、DTI、SWI序列,两种线圈)。
- 41例DBS患者行3-T MRI无不良事件。
临床怎么用
- 作者认为在本研究实验条件下,使用选定脉冲序列对DBS患者行3-T MRI似乎是安全的。
- 作者认为DBS患者一般应避免使用常规方案的T2加权序列。
- 作者提示限制性低于本研究患者所用条件的成像条件(如发射体部多阵列接收线圈)也可能安全,但该提示来自体模补充数据。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Physicians are more frequently encountering patients who are treated with deep brain stimulation (DBS), yet many MRI centers do not routinely perform MRI in this population. This warrants a safety assessment to improve DBS patients' accessibility to MRI, thereby improving their care while simultaneously providing a new tool for neuromodulation research. METHODS: A phantom simulating a patient with a DBS neuromodulation device (DBS lead model 3387 and IPG Activa PC model 37601) was constructed and used. Temperature changes at the most ventral DBS electrode contacts, implantable pulse generator (IPG) voltages, specific absorption rate (SAR), and B1+rms were recorded during 3-T MRI scanning. Safety data were acquired with a transmit body multi-array receive and quadrature transmit-receive head coil during various pulse sequences, using numerous DBS configurations from "the worst" to "the most common."In addition, 3-T MRI scanning (T1 and fMRI) was performed on 41 patients with fully internalized and active DBS using a quadrature transmit-receive head coil. MR images, neurological examination findings, and stability of the IPG impedances were assessed. RESULTS: In the phantom study, temperature rises at the DBS electrodes were less than 2°C for both coils during 3D SPGR, EPI, DTI, and SWI. Sequences with intense radiofrequency pulses such as T2-weighted sequences may cause higher heating (due to their higher SAR). The IPG did not power off and kept a constant firing rate, and its average voltage output was unchanged. The 41 DBS patients underwent 3-T MRI with no adverse event. CONCLUSIONS: Under the experimental conditions used in this study, 3-T MRI scanning of DBS patients with selected pulse sequences appears to be safe. Generally, T2-weighted sequences (using routine protocols) should be avoided in DBS patients. Complementary 3-T MRI phantom safety data suggest that imaging conditions that are less restrictive than those used in the patients in this study, such as using transmit body multi-array receive coils, may also be safe. Given the interplay between the implanted DBS neuromodulation device and the MRI system, these findings are specific to the experimental conditions in this study.