术中计算机断层扫描与透视在脑室腹腔分流管置入术中的比较
该回顾性分析纳入152例患者,比较术中计算机断层扫描(iCT,n=48)、双平面透视(n=57)与无术中成像对照组(n=47)的脑室导管置入准确性。iCT组导管位置准确24/48例,透视组45/57例(P=0.002),对照组23/47例;iCT未提高置管准确性,也未减少早期翻修。
要点
- iCT组导管准确置入24/48例,透视组45/57例(P=0.002),对照组23/47例。
- iCT组术后仍有4根导管位置异常,透视组为0根(P=0.03)。
- 作者认为透视可能是术中评估脑室导管位置的首选方法。
核心结论
- 作者认为透视可能是术中评估脑室导管位置的首选方法。
- 根据作者经验,iCT有更耗时的倾向,早期未伴随更陡的学习曲线。
- 另一考虑是每位患者的辐射暴露显著更高。
- iCT未改善导管置入准确性,也未减少脑室腹腔分流术患者的早期翻修。
研究设计与样本
- 回顾性分析纳入152例患者,导管置入通过iCT评估(n=48)和双平面透视评估(n=57)。
- 对照组(n=47)无术中成像。
- 结局指标包括脑室导管位置准确性、翻修手术和临床结局。
主要发现
- iCT组脑室导管置入准确24/48例,透视组45/57例(P=0.002),对照组23/47例。
- iCT估计最佳导管位置的敏感性和阳性预测值分别为100%和54%,特异性和阴性预测值分别为50%和100%。
- 术中翻修后,iCT组仍有4根导管位置异常,透视组为0根(P=0.03);这4根导管中2根术后翻修。
临床怎么用
- 作者认为透视可能是术中评估脑室导管位置的首选方法。
- 作者经验中iCT有更耗时的倾向,早期未伴随更陡的学习曲线,且每位患者辐射暴露显著更高。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Catheter malposition represents one of the major causes of ventriculoperitoneal (VP) shunt dysfunction. The usefulness of intraoperative fluoroscopy using skull landmarks has already been proved to decrease catheter malposition and surgical revision rates. After introducing intraoperative computed tomography (iCT) in our department, our objective was to evaluate the accuracy of this imaging modality to decrease cranial catheter misplacement compared with intraoperative fluoroscopy. METHODS: In our retrospective analysis of 152 patients, catheter placement was evaluated by iCT (n = 48) and biplane fluoroscopy (n = 57). A control group (n = 47) had no intraoperative imaging. Outcome measures included accuracy of ventricular catheter position, revision surgeries, and clinical outcomes. RESULTS: Ventricular catheter placement was accurate in 24/48 patients with iCT and 45/57 patients with fluoroscopy (P = 0.002) versus 23/47 patients in the control group. Sensitivity and positive predictive value for estimating optimal catheter position with iCT were 100% and 54%. The specificity and negative predictive value were 50% and 100%. After intraoperative revision, 4 catheters remained malpositioned in the iCT group, whereas the fluoroscopy group had none (P = 0.03); 2 of these 4 catheters were revised postoperatively. CONCLUSIONS: Fluoroscopy may be the method of choice to intraoperatively assess ventricular catheter positioning. In our experience, iCT shows a tendency to be more time consuming and, in the beginning, was not associated with a steeper learning curve. Another consideration was the significant higher radiation exposure per patient. iCT did not improve the accuracy of catheter placement and did not decrease early revisions for VP placement patients.
引用
- Wanderer S, Coluccia D, Añon J, et al. Intraoperative Computed Tomography Versus Fluoroscopy for Ventriculoperitoneal Shunt Placement. World Neurosurg. 2019. PMID: 30639491. doi:10.1016/j.wneu.2019.01.002