一例86岁右利手药物难治性特发性震颤患者接受左侧MRI引导聚焦超声丘脑毁损术,靶向齿状核红核丘脑束与丘脑腹中间核交界处,术后即刻症状改善且影像显示齿状核红核丘脑束中断。术后9周症状部分复发,MRI显示左侧丘脑消融灶回缩,遂于初次治疗后4个月行重复左侧MRgFUS丘脑毁损术,震颤减轻;第二次术中磁共振测温因初始手术的磁敏感效应而不可靠,最终以能量释放量监测超声发射。患者第二次术后恢复良好,12个月随访时震颤控制持续改善。
要点
- 重复MRgFUS丘脑毁损术可用于初次术后症状复发且影像显示消融灶回缩的特发性震颤患者。
- 第二次MRgFUS术中磁共振测温因初始手术的磁敏感效应出现大幅波动和假读数,测温不可靠。
- 最终超声发射以能量释放量进行监测,患者第二次术后恢复良好,12个月随访震颤控制持续改善。
核心结论
- 重复MRgFUS丘脑毁损术在本次病例中使患者震颤减轻,且12个月随访时震颤控制持续改善。
研究设计与样本
- 病例报告,描述一例86岁右利手药物难治性特发性震颤患者接受重复左侧MRgFUS丘脑毁损术并回顾系列MRI表现。
主要发现
- 初次左侧MRgFUS丘脑毁损术靶向齿状核红核丘脑束与丘脑腹中间核交界处,术后即刻症状改善,即刻术后影像显示齿状核红核丘脑束中断。
- 术后9周症状部分复发,MRI显示左侧丘脑消融灶回缩。
- 初次治疗后4个月行重复左侧MRgFUS丘脑毁损术,震颤减轻;第二次术中磁共振测温因初始手术的磁敏感效应出现大幅波动和假读数,测温不可靠,最终以能量释放量监测超声发射。
- 患者第二次术后恢复良好,12个月随访时震颤控制持续改善。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
An 86-year-old right-handed man with medically refractory essential tremor was treated using left-sided MRI-guided focused ultrasound (MRgFUS) thalamotomy targeting the dentatorubrothalamic tract (DRTT) at its intersection with the ventral intermediate nucleus of the thalamus, with immediate symptomatic improvement and immediate postprocedure imaging demonstrating disruption of the DRTT. The patient experienced a partial return of symptoms 9 weeks following the procedure, and MRI demonstrated retraction of the left thalamic ablation site. The patient underwent repeat left-sided MRgFUS thalamotomy 4 months after initial treatment, resulting in reduced tremor. MR thermometry temperature measurements during the second MRgFUS procedure were unreliable with large fluctuations and false readings, likely due to susceptibility effects from the initial MRgFUS procedure. Final sonications were therefore monitored using the amount of energy delivered. The patient fared well after the second procedure and had sustained improvement in tremor control at the 12-month follow-up. This is the first report to describe the technical challenges of repeat MRgFUS with serial imaging.
引用
- Weidman EK, Kaplitt MG, Strybing K, et al. Repeat magnetic resonance imaging-guided focused ultrasound thalamotomy for recurrent essential tremor: case report and review of MRI findings. J Neurosurg. 2019. PMID: 30684946. doi:10.3171/2018.10.JNS181721
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。