一种新型机器人引导的微创脑肿瘤活检技术
该研究提出一种新型微创机器人引导活检器械,并在连续66例立体定向活检中与标准钻孔活检比较。微创组32例、标准组34例,微创组在入点和靶点精度更高、切口至缝合时间更短、皮肤切口更短,所有病例均获得诊断性组织样本。
要点
- 微创机器人引导活检器械在全部患者中可行。
- 微创组入点精度中位数1.5 mm,标准组1.7 mm,p=0.008。
- 微创组靶点精度中位数1.5 mm,标准组2.0 mm,p=0.019。
核心结论
- 新型微创机器人引导活检器械在全部患者中可行,据作者数据可显著提高立体定向活检精度、缩短手术时间并改善美容效果。
研究设计与样本
- 连续66例机器人引导立体定向活检病例,比较微创组32例与标准钻孔组34例的可行性、精度、时长和结局。
主要发现
- 微创组入点精度中位数1.5 mm(范围0.2-3.2 mm)vs 标准组1.7 mm(范围0.8-5.1 mm),p=0.008。
- 微创组靶点精度中位数1.5 mm(范围0.4-3.4 mm)vs 标准组2.0 mm(范围0.8-3.9 mm),p=0.019。
- 微创组切口至缝合时间中位数30分钟(范围15-50分钟)vs 标准组37.5分钟(范围25-105分钟),p<0.001。
- 微创组皮肤切口中位数16.3 mm(范围12.7-23.4 mm)vs 标准组28.4 mm(范围20-42.2 mm),p=0.002。
- 全部病例均获得诊断性组织样本。
临床怎么用
- 作者认为,根据其数据,微创机器人引导操作可显著提高立体定向活检的精度、缩短手术时间并改善美容效果。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: As decisions regarding tumor diagnosis and subsequent treatment are increasingly based on molecular pathology, the frequency of brain biopsies is increasing. Robotic devices overcome limitations of frame-based and frameless techniques in terms of accuracy and usability. The aim of the present study was to present a novel, minimally invasive, robot-guided biopsy technique and compare the results with those of standard burr hole biopsy. METHODS: A tubular minimally invasive instrument set was custom-designed for the iSYS-1 robot-guided biopsies. Feasibility, accuracy, duration, and outcome were compared in a consecutive series of 66 cases of robot-guided stereotactic biopsies between the minimally invasive (32 patients) and standard (34 patients) procedures. RESULTS: Application of the minimally invasive instrument set was feasible in all patients. Compared with the standard burr hole technique, accuracy was significantly higher both at entry (median 1.5 mm [range 0.2-3.2 mm] vs 1.7 mm [range 0.8-5.1 mm], p = 0.008) and at target (median 1.5 mm [range 0.4-3.4 mm] vs 2.0 mm [range 0.8-3.9 mm], p = 0.019). The incision-to-suture time was significantly shorter (median 30 minutes [range 15-50 minutes] vs 37.5 minutes [range 25-105 minutes], p < 0.001). The skin incision was significantly shorter (median 16.3 mm [range 12.7-23.4 mm] vs 28.4 mm [range 20-42.2 mm], p = 0.002). A diagnostic tissue sample was obtained in all cases. CONCLUSIONS: Application of the novel instrument set was feasible in all patients. According to the authors' data, the minimally invasive robot-guidance procedure can significantly improve accuracy, reduce operating time, and improve the cosmetic result of stereotactic biopsies.
引用
- Minchev G, Kronreif G, Ptacek W, et al. A novel robot-guided minimally invasive technique for brain tumor biopsies. J Neurosurg. 2019. PMID: 30660122. doi:10.3171/2018.8.JNS182096
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。