深部脑刺激术中计算机断层扫描的实施:工作流程、精度与辐射暴露的陷阱与优化
本文报告18例运动障碍患者接受双侧定向电极深部脑刺激植入,术中采用32层移动CT联合微电极记录。术中CT虽多次扫描仍显著缩短总手术时间,低剂量方案使总有效剂量降至1.15 mSv,较标准非术中CT的DBS操作降低3.5倍;最终术中CT显示径向误差1.04±0.58 mm、矢量误差2.28±0.97 mm。
要点
- 术中CT可顺利整合入DBS手术流程,多次扫描下仍显著缩短总手术时间。
- 低剂量术中CT方案使总有效剂量降至1.15 mSv,较标准非术中CT的DBS操作降低3.5倍。
- 最终术中CT显示径向误差1.04±0.58 mm、矢量误差2.28±0.97 mm。
核心结论
- 术中CT可容易地整合入手术流程,形成整体高效且节省时间的手术。
- 反复术中扫描确保可靠的电极放置,低剂量扫描方案可防止过度的辐射暴露。
- 微电极的术中CT是可行的,并使1根电极得到调整。
研究设计与样本
- 18例患者接受双侧定向电极的DBS植入,采用32层移动CT扫描仪联合微电极记录。
主要发现
- 术中CT使总手术时间显著缩短,尽管进行了多次扫描。
- 在最初5例中的3例,术中CT导致最终电极的调整,体现了学习曲线及整合路径图以交换微电极至最终电极的必要性。
- 实施低剂量CT方案将微电极术中CT加入优化流程,导致1例患者术中调整轨迹。
- 低剂量方案使总有效剂量降至1.15 mSv,即较标准非术中CT的DBS操作降低3.5倍,尽管重复进行术中CT。
- 基于最终术中CT的术中电极检测显示,与术前计划相比(经微电极记录结果调整),径向误差为1.04±0.58 mm,矢量误差为2.28±0.97 mm。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Deep brain stimulation (DBS) is an effective treatment for movement disorders. Stereotactic electrode placement can be guided by intraoperative imaging, which also allows for immediate intraoperative quality control. This article is about implementation and refining a workflow applying intraoperative computed tomography (iCT) for DBS. METHODS: Eighteen patients underwent DBS with bilateral implantation of directional electrodes applying a 32-slice movable computed tomography scanner in combination with microelectrode recording. RESULTS: iCT led to a significant decrease in overall procedural time, despite performing multiple scans. In 3 of the initial 5 cases, iCT caused an adjustment of the final electrodes demonstrating the learning curve and the necessity to integrate road mapping for the exchange of microelectrode to final electrode. Implementation of low-dose computed tomography protocols added microelectrode iCT to the refined workflow, resulting in an intraoperative adjustment of a trajectory in 1 patient. Low-dose protocols lowered the total effective dose to 1.15 mSv, that is, a reduction by a factor of 3.5 compared to a standard non-iCT DBS procedure, despite repeated iCTs. Intraoperative lead detection based on final iCT revealed a radial error of 1.04 ± 0.58 mm and a vector error of 2.28 ± 0.97 mm compared to the preoperative planning, adjusted by the findings of microelectrode recording. CONCLUSIONS: iCT can be easily integrated into the surgical workflow resulting in an overall efficient time-saving procedure. Repeated intraoperative scanning ensures reliable electrode placement, although low-dose scanning protocols prevent extensive radiation exposure. iCT of microelectrodes is feasible and led to the adjustment of 1 electrode.
引用
- Carl B, Bopp M, Gjorgjevski M, et al. Implementation of Intraoperative Computed Tomography for Deep Brain Stimulation: Pitfalls and Optimization of Workflow, Accuracy, and Radiation Exposure. World Neurosurg. 2018. PMID: 30593970. doi:10.1016/j.wneu.2018.12.079
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。