3T立体定向磁共振成像的畸变校正方案:一项临床研究
这项前瞻性研究纳入20例拟行CT立体定向功能手术或脑内病灶活检的患者,以CT图像为参考计算MRI几何精度,采用正常与反转两种图像校正3T MRI几何畸变。畸变范围为0-5.6 mm,且随立体定向等中心距离增大而增大;校正后畸变最小0 mm、最大3.5 mm,CT与MRI校正后x坐标无显著差异(P > 0.05)。
要点
- 畸变范围为0-5.6 mm,随立体定向等中心距离增大而增大,周边更明显。
- 校正后畸变最小0 mm、最大3.5 mm。
- CT与MRI校正后x坐标无显著差异(P > 0.05)。
核心结论
- 所提出的方法可满意地校正临床3T MRI研究中的几何畸变。
- 在流程实现软件自动化后,该技术的临床使用可变得实用且高效。
- 该方法可应用于所有自旋回波MRI序列。
研究设计与样本
- 前瞻性研究,纳入20例拟行CT立体定向功能手术或脑内病灶活检的患者。
- 以CT图像为参考计算MRI几何精度,每次扫描获得正常与反转2幅图像。
- 在CT与MRI上选取8个不同点,共152个点,基于功效分析计算值>0.999;1例因无法在CT与MRI上找到可靠的共同标志点而被排除。
主要发现
- 畸变范围为0-5.6 mm,且随立体定向等中心距离增大而成比例增大,即周边畸变更大。
- 校正后所发现的最小和最大畸变分别为0 mm和3.5 mm。
- CT与MRI校正后x坐标之间无显著差异(P > 0.05)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: With application of 3T magnetic resonance imaging (MRI) to functional neurosurgery procedures and given the inherent requirement of millimetric precision, the need to develop a method for correction of geometric image distortion emerged. The aim of this study was to demonstrate clinical safety and practical viability of a correction protocol in patients scheduled to undergo stereotactic procedures using 3T MRI. METHODS: This prospective study comprised 20 patients scheduled to undergo computed tomography (CT) stereotactic functional procedures or encephalic brain lesion biopsies. The CT images were references for MRI geometric accuracy calculations. For each scan, 2 images were obtained: normal and reversed images. Eight distinct points on CT and MRI were selected summing 152 points that were based on a power analysis calculation value >0.999. One patient was excluded because of the inability to find reliable common landmark points on CT and MRI. RESULTS: The distortion range was 0-5.6 mm and increased proportionally with stereotactic isocenter distance, meaning the distortion was greater in the periphery. After correction, the minimum and maximum distortion found was 0 mm and 3.5 mm, respectively. There was no significant difference between CT and MRI corrected x-coordinates (P > 0.05). CONCLUSIONS: The proposed method can satisfactorily correct geometric distortions in clinical 3T MRI studies. Clinical use of the technique can be practical and efficient after software automation of the process. The method can be applied to all spin-echo MRI sequences.
引用
- Tavares WM, Tustumi F, da Costa Leite C, et al. Distortion Correction Protocol for 3T Stereotactic Magnetic Resonance Imaging: A Clinical Study. World Neurosurg. 2018. PMID: 30394358. doi:10.1016/j.wneu.2018.10.123
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。