Enterprise与Pipeline支架血栓栓塞事件比较:猪体内实验
该研究在6只约克夏猪颈动脉分叉处对侧植入尺寸匹配的Enterprise和Pipeline栓塞装置,于基线及2周随访时行血管造影并测量迷网灌注面积;无血栓栓塞事件时灌注迷网面积为134.4±6.9 mm²,发生事件时降至37.1±8.6 mm²(P<0.01);两种装置血栓栓塞事件率均为33.3%(P>0.99),血栓性无差异。
要点
- 猪颈动脉分叉支架模型可用于体内评估装置相关血栓性和远端栓塞事件率。
- 灌注迷网面积缩小提示血栓栓塞事件发生:事件组37.1±8.6 mm² vs 无事件组134.4±6.9 mm²(P<0.01)。
- Enterprise与Pipeline血栓栓塞事件率均为33.3%,两种装置血栓性相似(P>0.99)。
核心结论
- 猪颈动脉支架模型允许体内分析装置相关血栓性和远端栓塞事件率。
- 灌注迷网面积缩小提示血栓栓塞事件的存在。
- 在本实验中,Enterprise与Pipeline相关血栓性相似。
研究设计与样本
- 采用6只约克夏猪,在颈动脉分叉处对侧植入尺寸匹配的装置,每只动物接受1枚Enterprise和1枚PED。
- 行基线和2周随访血管造影,分析图像包括对迷网(由颈总动脉分支供血的小血管团)的详细测量。
主要发现
- 2周时无血栓栓塞事件者(n=8),灌注迷网面积为134.4±6.9 mm²。
- 存在血栓栓塞事件者(n=4),灌注迷网面积缩小至37.1±8.6 mm²(P<0.01)。
- Enterprise与PED之间血栓性无差异,两种装置血栓栓塞事件率均为33.3%(P>0.99)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Thromboembolic complications are serious adverse events associated with stenting of intracranial aneurysms. For the development of novel devices, currently no practical model exists to evaluate distal embolic events. The purpose of the current study was 2-fold: 1) assess the feasibility of a porcine carotid bifurcation model for in vivo testing of stent thrombogenicity and 2) compare the thrombogenicity of the Enterprise vascular reconstruction device versus the Pipeline embolization device (PED). METHODS: Six Yorkshire pigs underwent implantation of size-matched devices across the carotid bifurcation. Each animal received 1 Enterprise and 1 PED contralaterally. Baseline and 2-week follow-up angiography were performed. Images were analyzed including detailed measurements of the rete mirabile, a small-vessel conglomerate supplied by branches of the common carotid artery. RESULTS: In the absence of thromboembolic events at 2 weeks (n = 8), the perfused rete area measured 134.4 ± 6.9 mm2. In the presence of thromboembolic events (n = 4), the perfused rete was reduced to an area covering 37.1 ± 8.6 mm2 (P < 0.01). There was no difference in thrombogenicity between Enterprise and PED with a thromboembolic event rate of 33.3% for both devices (P > 0.99). CONCLUSION: The swine carotid stent model allows in vivo analysis of device-related thrombogenicity and distal embolic event rates. Size reduction of the perfused rete mirabile indicates the presence of thromboembolic events. Both Enterprise and PED were associated with similar thrombogenicity in this experiment.
引用
- Griffin A, Chaparro E, Fedorova E, et al. Thromboembolic Events with Enterprise Versus Pipeline: Porcine In Vivo Experiment. World Neurosurg. 2018. PMID: 30468928. doi:10.1016/j.wneu.2018.11.073
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。