动静脉瘘栓塞术中微导管意外破裂后Pipeline支架装置的新用途

Novel Use of Pipeline Stent Device After Inadvertent Microcatheter Rupture During Arteriovenous Fistula Embolization.
PubMed 文献 · World Neurosurg · 2018-07-30

本文报告一例破裂动静脉瘘患者在栓塞过程中发生微导管近端破裂及Onyx溢出,通过在椎动脉V3段部署Pipeline栓塞装置将Onyx限制于导管与血管壁之间,并辅以球囊血管成形术完成支架打开,手术顺利,随访造影显示后循环通畅及6个月支架通畅。作者认为微导管破裂和Onyx溢出是潜在并发症,处理应个体化,PED的独特闭环设计成功避免了可能危及生命的椎基底动脉系统闭塞。

要点

  • 微导管破裂伴Onyx溢出是动静脉畸形/动静脉瘘栓塞的潜在并发症
  • PED成功将溢出的Onyx限制于椎动脉V3段导管与血管壁之间
  • 6个月随访造影显示支架通畅

核心结论

  • 微导管破裂和Onyx溢出是动静脉畸形/动静脉瘘栓塞的潜在并发症,处理应个体化。
  • PED的独特闭环设计成功用于避免可能危及生命的椎基底动脉系统闭塞。
  • 据作者所知,这是首次报告PED的这种新用途。

研究设计与样本

  • 病例报告:一例破裂动静脉瘘患者在栓塞过程中发生微导管近端破裂及Onyx溢出。

主要发现

  • 在椎动脉第三段(V3段)部署Pipeline栓塞装置(PED),将Onyx限制于导管与血管壁之间。
  • 球囊血管成形术帮助完成支架打开。
  • 手术顺利,患者恢复良好。
  • 随访造影显示后循环通畅;6个月随访造影显示支架通畅。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Onyx, a liquid embolic agent, is the mainstay of embolization treatment of arteriovenous malformation or arteriovenous fistula. Microcatheter retention in an embolic cast is a well-known complication. Rupture of catheter with spillage of onyx is a rare phenomenon but can lead to potential occlusion of the parent vessel. METHODS: We describe a case in which a patient with ruptured arteriovenous fistula experienced rupture of a microcatheter at the proximal part and spillage of onyx on embolization. A Pipeline embolization device (PED) was successfully deployed, and onyx was jailed between the catheter and vessel wall at the third segment of the vertebral artery (V3 segment). A balloon angioplasty helped to complete the opening of the stent. RESULTS: The procedure was uneventful, and the patient recovered well. Follow-up angiogram revealed patency of posterior circulation. The 6-month follow-up angiogram revealed a patent stent. CONCLUSIONS: Microcatheter rupture and spillage of onyx during embolization of arteriovenous malformation/arteriovenous fistula is a potential complication, and management should be individualized. The unique close cell design of the PED was successfully used to avoid a potentially life-threatening occlusion of the vertebrobasilar system. To the best of our knowledge, we report for the first time this novel use of PED.

来源:World Neurosurg,查看原文。版权归原作者所有。