脊柱手术中Br(E)-MsEP监测联合波形波幅与起始潜伏期的新报警标准
单中心回顾性研究纳入83例脊柱手术患者,术中行Br(E)-MsEP监测,分析术后运动障碍与起始潜伏期及波幅下降≥70%联合标准的关系。结果显示联合标准预测术后运动障碍的敏感度100%、特异度93%。
要点
- 起始潜伏期较基线延迟15%预测术后运动障碍的敏感度78%、特异度96%
- 波幅下降≥70%联合起始潜伏期延迟≥10%的敏感度100%、特异度93%
- 联合标准假阳性率7%、假阴性率0%、阳性预测值64%、阴性预测值100%
核心结论
- 在术中Br(E)-MsEP脊髓监测中,联合波幅下降≥70%与起始潜伏期延迟≥10%的报警标准具有高特异度,可减少假阳性结果。
研究设计与样本
- 单中心回顾性研究,纳入83例使用术中Br(E)-MsEP监测的脊柱手术患者,共选取四肢1726块肌肉进行监测,其中1640块(95%)获得可接受的基线Br(E)-MsEP反应。
主要发现
- 83例患者中9例出现术后运动障碍。
- 有术后运动障碍与无术后运动障碍患者的起始潜伏期延迟相比差异显著(1.09%±0.06% vs 1.31%±0.14%,p<0.01)。
- ROC分析中,术中起始潜伏期较基线延迟15%预测术后运动障碍的敏感度78%、特异度96%。
- 进一步ROC分析中,波幅下降≥70%联合起始潜伏期延迟≥10%预测术后运动障碍的敏感度100%、特异度93%、假阳性率7%、假阴性率0%、阳性预测值64%、阴性预测值100%。
关键数值与推荐
- 波幅下降≥70%联合起始潜伏期延迟≥10%的报警标准:敏感度100%、特异度93%、假阳性率7%、假阴性率0%、阳性预测值64%、阴性预测值100%。
- 起始潜伏期延迟15%的报警标准:敏感度78%、特异度96%。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Monitoring of brain evoked muscle-action potentials (Br[E]-MsEPs) is a sensitive method that provides accurate periodic assessment of neurological status. However, occasionally this method gives a relatively high rate of false-positives, and thus hinders surgery. The alarm point is often defined based on a particular decrease in amplitude of a Br(E)-MsEP waveform, but waveform latency has not been widely examined. The purpose of this study was to evaluate onset latency in Br(E)-MsEP monitoring in spinal surgery and to examine the efficacy of an alarm point using a combination of amplitude and latency. METHODS: A single-center, retrospective study was performed in 83 patients who underwent spine surgery using intraoperative Br(E)-MsEP monitoring. A total of 1726 muscles in extremities were chosen for monitoring, and acceptable baseline Br(E)-MsEP responses were obtained from 1640 (95%). Onset latency was defined as the period from stimulation until the waveform was detected. Relationships of postoperative motor deficit with onset latency alone and in combination with a decrease in amplitude of ≥ 70% from baseline were examined. RESULTS: Nine of the 83 patients had postoperative motor deficits. The delay of onset latency compared to the control waveform differed significantly between patients with and without these deficits (1.09% ± 0.06% vs 1.31% ± 0.14%, p < 0.01). In ROC analysis, an intraoperative 15% delay in latency from baseline had a sensitivity of 78% and a specificity of 96% for prediction of postoperative motor deficit. In further ROC analysis, a combination of a decrease in amplitude of ≥ 70% and delay of onset latency of ≥ 10% from baseline had sensitivity of 100%, specificity of 93%, a false positive rate of 7%, a false negative rate of 0%, a positive predictive value of 64%, and a negative predictive value of 100% for this prediction. CONCLUSIONS: In spinal cord monitoring with intraoperative Br(E)-MsEP, an alarm point using a decrease in amplitude of ≥ 70% and delay in onset latency of ≥ 10% from baseline has high specificity that reduces false positive results.
引用
- Kobayashi K, Ando K, Shinjo R, et al. A new criterion for the alarm point using a combination of waveform amplitude and onset latency in Br(E)-MsEP monitoring in spine surgery. J Neurosurg Spine. 2018. PMID: 30052151. doi:10.3171/2018.3.SPINE171348
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。