猪模型中的机器人辅助MRI引导介入性间质磁共振引导聚焦超声消融
该研究在猪模型中使用磁共振引导的机器人辅助系统进行间质聚焦超声消融,先优化机器人在磁共振室的工作流程并改装机械臂,随后对6头农场猪实施7次消融,其中2头为存活手术,术后2周无神经后遗症,术中及术后影像与组织学检查相关。
要点
- 机器人辅助间质聚焦超声消融在猪脑内成功实施,术后2周无神经后遗症。
- 术后即刻弥散加权成像和T1加权磁共振准确反映消融体积。
- 术后1周T2和液体衰减反转恢复序列图像准确反映消融体积。
核心结论
- 研究成功在猪中实施磁共振引导机器人辅助间质聚焦超声消融,术中及术后影像与组织学检查相关。
研究设计与样本
- 动物实验:先对3只动物优化机器人在磁共振室的工作流程并改装机械臂以扩大活动范围,随后对6头农场猪(4头急性、2头存活)实施7次间质磁共振引导聚焦超声消融。
主要发现
- 2头存活手术猪在术后2周无神经后遗症。
- 间质聚焦超声治疗后即刻,弥散加权成像和T1加权磁共振准确反映消融体积。
- 术后1周,T2和液体衰减反转恢复序列图像准确反映消融体积。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Ablative lesions are current treatments for epilepsy and brain tumors. Interstitial magnetic resonance (MR) guided focused ultrasound (iMRgFUS) may be an alternate ablation technique which limits thermal tissue charring as compared to laser therapy (LITT) and can produce larger ablation patterns nearer the surface than transcranial MR guided focused ultrasound (tcMRgFUS). OBJECTIVE: To describe our experience with interstitial focused ultrasound (iFUS) ablations in swine, using MR-guided robotically assisted (MRgRA) delivery. METHODS: In an initial 3 animals, we optimized the workflow of the robot in the MR suite and made modifications to the robotic arm to allow range of motion. Then, 6 farm pigs (4 acute, 2 survival) underwent 7 iMRgFUS ablations using MRgRA. We altered dosing to explore differences between thermal dosing in brain as compared to other tissues. Imaging was compared to gross examination. RESULTS: Our work culminated in adjustments to the MRgRA, iMRgFUS probes, and dosing, culminating in 2 survival surgeries; swine had ablations with no neurological sequelae at 2 wk postprocedure. Immediately following iMRgFUS therapy, diffusion-weighted imaging, and T1 weighted MR were accurate reflections of the ablation volume. T2 and fluid-attenuated inversion-recovery (FLAIR) images were accurate reflections of ablation volume 1-wk postprocedure. CONCLUSION: We successfully performed MRgRA iFUS ablation in swine and found intraoperative and postoperative imaging to correlate with histological examination. These data are useful to validate our system and to guide imaging follow-up for thermal ablation lesions in brain tissue from our therapy, tcMRgFUS, and LITT.
引用
- MacDonell J, Patel N, Fischer G, et al. Robotic Assisted MRI-Guided Interventional Interstitial MR-Guided Focused Ultrasound Ablation in a Swine Model. Neurosurgery. 2019. PMID: 29905844. doi:10.1093/neuros/nyy266
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。