SiLuDrain 试验:比较标准腰椎引流管与银浸渍腰椎引流管的前瞻性随机对照试验
该前瞻性随机对照试验将 48 例需放置外部腰椎引流(ELD)的患者随机分配至常规 ELD 或银浸渍 ELD(si-ELD)组,评估脑脊液感染与器械相关并发症。si-ELD 组感染率 4.2% vs 常规组 16.7%(p=0.16),器械相关并发症 8.3% vs 37.5%(p=0.02),可用性评分 si-ELD 显著更优(p=0.003)。作者认为 si-ELD 在操作性和器械相关并发症方面表现更佳,但是否优于抗生素浸渍导管或抗生素预防方案尚待证实。
要点
- si-ELD 组脑脊液感染率 4.2% vs 常规组 16.7%,差异无统计学意义(p=0.16)。
- si-ELD 组器械相关并发症 8.3% vs 常规组 37.5%,差异有统计学意义(p=0.02)。
- si-ELD 可用性评分显著优于常规 ELD(p=0.003)。
核心结论
- 银浸渍 ELD 导管可能减少脑脊液感染,在操作性和器械相关并发症方面表现显著更优。
- 银浸渍 ELD 是否优于抗生素浸渍导管或包含抗生素预防的临床方案尚待证实。
研究设计与样本
- 前瞻性随机对照试验,将需放置 ELD 的患者随机分配至常规 ELD 或银浸渍 ELD 导管组。
- 共纳入 48 例患者,定期评估脑脊液感染与器械相关并发症。
- 放置 ELD 的神经外科医生采用 6 项序贯性能量表(范围 1 分极差至 5 分极好)评价导管可用性与操作性。
- 对研究中分离的所有微生物采用导管滚动法检测银敏感性。
主要发现
- si-ELD 组感染率 4.2% vs 常规组 16.7%,差异无统计学意义(p=0.16)。
- 大多数感染由葡萄球菌属引起。
- si-ELD 组器械相关并发症 8.3% vs 常规组 37.5%,差异有统计学意义(p=0.02)。
- si-ELD 可用性评分显著优于常规 ELD(p=0.003)。
- si-ELD 对多种葡萄球菌属显示抗菌活性,但近平滑念珠菌和大肠埃希菌不受该抗菌剂影响。
临床怎么用
- 作者认为银浸渍 ELD 导管可能减少脑脊液感染,在操作性和器械相关并发症方面表现显著更优。
- 作者指出银浸渍 ELD 是否优于抗生素浸渍导管或包含抗生素预防的临床方案尚待证实。
局限性
- 银浸渍 ELD 是否优于抗生素浸渍导管或包含抗生素预防的临床方案尚待证实。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Catheter-associated cerebrospinal fluid (CSF) infection remains a serious event, especially for patients in neurocritical care units. The use of external ventricular drain (EVD) catheters impregnated with antimicrobial substances has led to a significant reduction of infection rates. This study was undertaken to compare the use of antimicrobial, silver-impregnated external lumbar drains (si-ELDs) and conventional ELDs. METHODS: Patients with an indication for ELD placement were randomized to receive either a conventional or an si-ELD catheter. Regular assessment for CSF infections and device-related complications was performed. Neurosurgeons placing the ELD rated the usability and handling of the catheter on a 6-item ordinal performance scale (range: 1, very bad, to 5, very good). All microorganisms isolated in this study were tested for silver-susceptibility via a catheter-roll method. RESULTS: A total of 48 patients were enrolled in the trial. The si-ELD catheters showed a nonsignificantly lower infection rate compared to conventional ones (4.2% vs 16.7%, p = 0.16). The majority of infections were caused by Staphylococcus species. Device-related complications occurred significantly less often with silver-impregnated-catheters than with conventional ones (8.3% vs 37.5%, p = 0.02). The usability was rated significantly better for si-ELDs (p = 0.003). Antimicrobial susceptibility was shown for si-ELDs against various Staphylococcus spp., but Candida parapsilosis and Escherichia coli were not affected by this antimicrobial agent. CONCLUSIONS: Silver-impregnated ELD catheters, which could potentially reduce the number of CSF infections, show significantly better properties in regard to handling and fewer device-related complications. Whether they are superior to antibiotic-impregnated catheters or a clinical regimen involving antibiotic prophylaxis remains to be proven.Clinical trial registration no.: DRKS00013513 (Deutsches Register Klinischer Studien).
引用
- Jakobs M, Klein S, Eigenbrod T, et al. The SiLuDrain Trial: a prospective randomized controlled trial comparing standard versus silver-impregnated lumbar drains. J Neurosurg. 2018. PMID: 29905509. doi:10.3171/2018.1.JNS181522
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。