EOS低剂量双平面X线用于脑积水儿童分流管系列检查的初步研究
本研究回顾性分析9例(20岁以下)脑室腹腔分流管患儿,在6个月内因骨科指征接受EOS低剂量双平面X线检查,评估分流管可见性与完整性。所有患者分流管完整性在EOS图像上均易显示,2例分流管断开经EOS确认,未遗漏任何分流管相关信息。作者认为EOS可作为替代常规X线分流管系列检查的技术。
要点
- 9例脑室腹腔分流管患儿接受EOS低剂量双平面X线检查,分流管完整性均易显示。
- 2例分流管断开经EOS图像确认,与其它X线图像相比未遗漏分流管信息。
- EOS无需多次辐射暴露和图像拼接处理,检查时间更短且辐射显著减少。
核心结论
- 这些初步结果提示EOS可作为替代常规X线分流管系列检查的技术来显示分流管完整性。
研究设计与样本
- 回顾性研究,在6个月期间本中心因各种骨科指征(主要为脊柱侧弯评估)行EOS检查的9例20岁以下脑室腹腔分流管患儿作为研究组。
主要发现
- 3例患者为双侧分流管,8例有既往X线片可比对。
- 所有患者分流管完整性在EOS图像上均易显示。
- 2例患者分流管断开经EOS图像确认。
- 与其它X线图像相比,EOS未遗漏任何分流管相关信息。
临床怎么用
- 作者认为EOS可能作为替代常规X线分流管系列检查的技术来显示分流管完整性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Shunt series (SS) are a common diagnostic tool used to verify shunt integrity. SS include X-ray films of the skull, chest, and abdomen and often are performed either when a shunted patient presents with suspected shunt malfunction or as a screening test to identify shunt disconnections or dislodgment. EOS low-dose biplanar X-rays are associated with significantly reduced radiation doses compared with ordinary X-rays and are used for various indications. This is the first publication on the use of EOS as a SS technique. METHODS: Over a period of 6 months, EOS were performed at our center for various orthopedic indications, mostly for scoliosis evaluation. Nine children (<20 years of age) had a ventriculoperitoneal shunt and served as the study group. We retrospectively reviewed shunt visibility and integrity in the EOS scans as well as regular SS or plain spinal X-rays. RESULTS: Three patients had bilateral shunts, and 8 had previous X-rays for comparison. In all patients, the shunt integrity was easily demonstrated on the EOS images. Two patients had an identified shunt disconnection confirmed on the EOS images. No shunt-related information was missed on the EOS compared with the other X-ray images. CONCLUSIONS: These preliminary results suggest that EOS may be used as an alternative technology to demonstrate shunt integrity instead of regular X-ray SS. Favorable shunt visibility without the need for multiple radiation exposures and image processing (such as stitching) results in a significantly shorter examination time and significant less radiation.
引用
- Ben-Sira L, Shiran SI, Pratt LT, et al. Use of EOS Low-Dose Biplanar X-Ray for Shunt Series in Children with Hydrocephalus: A Preliminary Study. World Neurosurg. 2018. PMID: 29730101. doi:10.1016/j.wneu.2018.04.187
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。