文献速递功能神经外科

MR引导聚焦超声脑功能手术治疗所致颅骨骨髓损伤

Skull bone marrow injury caused by MR-guided focused ultrasound for cerebral functional procedures.
PubMed 文献 · J Neurosurg · 2018-05-04

本文回顾性分析前30例接受MR引导聚焦超声(MRgFUS)治疗的脑功能疾病患者(1例尝试苍白球毁损术、29例丘脑腹中间核毁损术),发现7例出现新发颅骨病灶,所有患者均完全无症状;有颅骨病灶者应用的最大能量显著高于无病灶者,平均颅骨密度比(SDR)与病灶有无无相关性。作者认为低SDR区域可能因局部产热足以造成骨损伤,建议未来研究病灶位置与局部颅骨特性的相关性。

要点

  • 30例MRgFUS治疗患者中7例出现新发颅骨病灶,均完全无症状。
  • 有颅骨病灶者Exablate系统应用的最大能量显著高于无病灶者。
  • 平均颅骨密度比(SDR)与颅骨病灶有无无相关性。

核心结论

  • 局部颅骨密度、厚度和SDR因部位而异,低SDR区域可能因能量传递足以产生局部加热而形成骨病灶;未来研究应分析病灶位置与局部颅骨特性的相关性。

研究设计与样本

  • 回顾性分析前30例接受MRgFUS治疗的脑功能疾病患者,其中1例尝试苍白球毁损术、29例丘脑腹中间核毁损术,检查平均SDR与产生或尝试产生脑病灶所用最大能量的相关性。

主要发现

  • 30例接受MRgFUS治疗运动障碍的患者中,7例发现新发颅骨病灶,这些病灶在治疗前不存在且治疗后第1天MRI不可见。
  • 部分有或无颅骨病灶的患者在治疗时报告不适;所有有颅骨病灶的患者完全无症状。
  • 平均SDR与颅骨病灶有无无相关性,但有颅骨病灶者应用Exablate系统的最大能量显著高于无颅骨病灶者。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: One patient for whom an MR-guided focused ultrasound (MRgFUS) pallidotomy was attempted was discovered to have multiple new skull lesions with the appearance of infarcts on the MRI scan 3 months after his attempted treatment. The authors conducted a retrospective review of the first 30 patients treated with MRgFUS to determine the incidence of skull lesions in patients undergoing these procedures and to consider possible causes. METHODS: A retrospective review of the MRI scans of the first 30 patients, 1 attempted pallidotomy and 29 ventral intermediate nucleus thalamotomies, was conducted. The correlation of the mean skull density ratio (SDR) and the maximum energy applied in the production or attempted production of a brain lesion was examined. RESULTS: Of 30 patients treated with MRgFUS for movement disorders, 7 were found to have new skull lesions that were not present prior to treatment and not visible on the posttreatment day 1 MRI scan. Discomfort was reported at the time of treatment by some patients with and without skull lesions. All patients with skull lesions were completely asymptomatic. There was no correlation between the mean SDR and the presence or absence of skull lesions, but the maximum energy applied with the Exablate system was significantly greater in patients with skull lesions than in those without. CONCLUSIONS: It is known that local skull density, thickness, and SDR vary from location to location. Sufficient energy transfer resulting in local heating sufficient to produce a bone lesion may occur in regions of low SDR. A correlation of lesion location and local skull properties should be made in future studies.

来源:J Neurosurg,查看原文。版权归原作者所有。

引用

  1. Schwartz ML, Yeung R, Huang Y, et al. Skull bone marrow injury caused by MR-guided focused ultrasound for cerebral functional procedures. J Neurosurg. 2018. PMID: 29726769. doi:10.3171/2017.11.JNS17968

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。