一种新型神经假体实现长期皮层脑电记录与皮层刺激用于癫痫及脑机接口
本研究在非人灵长类动物(食蟹猴)左侧感觉运动皮层植入完全可植入且体外可充电的无线神经假体(ECOGIW-16E,含16触点硬膜下栅格),进行为期6个月的无线皮层脑电(ECoG)记录和直接皮层刺激(DCS)。结果显示植入耐受良好,信号质量优异,平均阻抗约40 KΩ,组织学仅见轻度炎症和单个显微镜下皮层坏死点。
要点
- 无线神经假体在非人灵长类中实现长达6个月的ECoG记录和多次DCS,耐受良好无后遗症。
- 平均电极阻抗约40 KΩ,远低于100 KΩ阈值,信号质量保持优异。
- 组织学显示栅格下皮层无肉眼可见损伤,仅电刺激区域见单个显微镜下坏死点及轻度炎症。
核心结论
- 该初步非人灵长类实验表明,无线神经假体实现长达6个月的长期ECoG记录和多次DCS,耐受良好且无后遗症。
- 作者预测癫痫手术可从这种新型假体中获益,因其可延长ECoG记录时间。
研究设计与样本
- 在非人灵长类(食蟹猴)左侧感觉运动皮层植入ECOGIW-16E装置(含16触点硬膜下栅格),记录ECoG信号6个月。
- 每日测量电极阻抗,目标维持阻抗值低于100 KΩ阈值。
- 在第1、3、6个月通过无线皮层刺激进行脑功能定位。
- 6个月后取出装置,采用常规组织学和免疫组织学检查评估长期植入后是否有损伤证据。
主要发现
- 植入耐受良好,猴未出现神经或行为学后果,术后数小时内恢复正常活动。
- 无线ECoG信号质量在6个月观察期内保持优异。
- 阻抗值远低于阈值,每个触点平均阻抗约40 KΩ。
- 无线皮层刺激可诱导上下肢运动,并引发手指精细运动。
- 取出时栅格被新形成的硬膜片包裹,取出容易,栅格与皮层无直接粘连。
临床怎么用
- 原文未据此给出临床推荐。
- 作者预测癫痫手术可从这种新型假体中获益,因其可延长ECoG记录时间。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Wireless technology is a novel tool for the transmission of cortical signals. Wireless electrocorticography (ECoG) aims to improve the safety and diagnostic gain of procedures requiring invasive localization of seizure foci and also to provide long-term recording of brain activity for brain-computer interfaces (BCIs). However, no wireless devices aimed at these clinical applications are currently available. The authors present the application of a fully implantable and externally rechargeable neural prosthesis providing wireless ECoG recording and direct cortical stimulation (DCS). Prolonged wireless ECoG monitoring was tested in nonhuman primates by using a custom-made device (the ECoG implantable wireless 16-electrode [ECOGIW-16E] device) containing a 16-contact subdural grid. This is a preliminary step toward large-scale, long-term wireless ECoG recording in humans. METHODS: The authors implanted the ECOGIW-16E device over the left sensorimotor cortex of a nonhuman primate (Macaca fascicularis), recording ECoG signals over a time span of 6 months. Daily electrode impedances were measured, aiming to maintain the impedance values below a threshold of 100 KΩ. Brain mapping was obtained through wireless cortical stimulation at fixed intervals (1, 3, and 6 months). After 6 months, the device was removed. The authors analyzed cortical tissues by using conventional histological and immunohistological investigation to assess whether there was evidence of damage after the long-term implantation of the grid. RESULTS: The implant was well tolerated; no neurological or behavioral consequences were reported in the monkey, which resumed his normal activities within a few hours of the procedure. The signal quality of wireless ECoG remained excellent over the 6-month observation period. Impedance values remained well below the threshold value; the average impedance per contact remains approximately 40 KΩ. Wireless cortical stimulation induced movements of the upper and lower limbs, and elicited fine movements of the digits as well. After the monkey was euthanized, the grid was found to be encapsulated by a newly formed dural sheet. The grid removal was performed easily, and no direct adhesions of the grid to the cortex were found. Conventional histological studies showed no cortical damage in the brain region covered by the grid, except for a single microscopic spot of cortical necrosis (not visible to the naked eye) in a region that had undergone repeated procedures of electrical stimulation. Immunohistological studies of the cortex underlying the grid showed a mild inflammatory process. CONCLUSIONS: This preliminary experience in a nonhuman primate shows that a wireless neuroprosthesis, with related long-term ECoG recording (up to 6 months) and multiple DCSs, was tolerated without sequelae. The authors predict that epilepsy surgery could realize great benefit from this novel prosthesis, providing an extended time span for ECoG recording.
引用
- Romanelli P, Piangerelli M, Ratel D, et al. A novel neural prosthesis providing long-term electrocorticography recording and cortical stimulation for epilepsy and brain-computer interface. J Neurosurg. 2018. PMID: 29749917. doi:10.3171/2017.10.JNS17400
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。