难治性癫痫患者丘脑前核的手术入路:国际多中心注册研究(MORE)的经验
基于MORE注册研究,在17个欧洲中心的73例丘脑前核脑深部电刺激植入中,比较经脑室与经脑室外导线轨迹将触点置于丘脑前核靶区域的成功率;经脑室轨迹约90%至少1个触点位于靶区域,经脑室外轨迹为71%;双侧至少1个触点位于靶区域的成功率经脑室植入为84%,经脑室外植入为58%(P < .05;Fisher精确检验);未观察到颅内出血,经脑室外导线轨迹后报告1例皮质梗死。
要点
- 经脑室导线轨迹约90%至少1个触点位于丘脑前核靶区域,经脑室外轨迹为71%。
- 双侧至少1个触点位于靶区域的成功率经脑室植入84% vs 经脑室外植入58%(P < .05;Fisher精确检验)。
- 未观察到颅内出血,经脑室外导线轨迹后报告1例皮质梗死。
核心结论
- 该注册研究结果支持丘脑前核脑深部电刺激采用经脑室导线轨迹,因其将触点置于丘脑前核靶区域的概率更高,且似乎未影响手术安全性。
- MORE注册研究的随访数据收集仍在继续,这些数据将提供与经脑室和经脑室外轨迹相关的结局。
研究设计与样本
- MORE(Medtronic Registry for Epilepsy)为开放标签观察性研究,评估丘脑前核脑深部电刺激治疗难治性癫痫的长期有效性、安全性和性能。
- 比较参与MORE注册研究的17个欧洲中心73例丘脑前核脑深部电刺激植入中经脑室与经脑室外导线轨迹将触点置于丘脑前核靶区域的成功率差异。
- 采用结合个体患者影像信息和立体定向图谱信息的筛选方法识别位于丘脑前核靶区域的触点。
主要发现
- 经脑室外导线轨迹占53%的轨迹。
- 经脑室导线轨迹约90%至少1个触点位于丘脑前核靶区域,经脑室外导线轨迹仅71%。
- 双侧至少1个触点位于丘脑前核靶区域的成功率经脑室植入为84%,经脑室外植入为58%(P < .05;Fisher精确检验)。
- 未观察到颅内出血,但经脑室外导线轨迹后报告1例皮质梗死。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The Medtronic Registry for Epilepsy (MORE; Medtronic Inc, Dublin, Ireland) is an open label observational study evaluating the long-term effectiveness, safety, and performance of deep brain stimulation (DBS) of the anterior nucleus of thalamus (ANT) for the treatment of refractory epilepsy. OBJECTIVE: To compare the difference in success rate of placing contacts at ANT-target region (ANT-TR) between transventricular (TV) and extraventricular (EV) lead trajectories in 73 ANT-DBS implants in 17 European centers participating in the MORE registry. METHODS: The success rate of placing contacts at ANT-TR was evaluated using a screening method combining both individual patient imaging information and stereotactic atlas information to identify contacts at ANT-TR. RESULTS: EV lead trajectory was used in 53% of the trajectories. Approximately, 90% of the TV lead trajectories had at least 1 contact at ANT-TR, vs only 71% of the EV lead trajectories. The success rate for placing at least 1 contact at ANT-TR bilaterally was 84% for TV implants and 58% for EV implants (P < .05; Fisher's exact). No intracranial bleedings were observed, but 1 cortical infarct was reported following EV lead trajectory. CONCLUSION: The results of this registry support the use of TV lead trajectories for ANT-DBS as they have a higher probability in placing contacts at ANT-TR, without appearing to compromise procedural safety. Follow-up data collection is continuing in the MORE registry. These data will provide outcomes associated with TV and EV trajectories.
引用
- Lehtimäki K, Coenen VA, Gonçalves Ferreira A, et al. The Surgical Approach to the Anterior Nucleus of Thalamus in Patients With Refractory Epilepsy: Experience from the International Multicenter Registry (MORE). Neurosurgery. 2019. PMID: 29554309. doi:10.1093/neuros/nyy023
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。