脑深部电刺激脉冲发生器更换手术的并发症:单中心回顾性研究
该单中心回顾性研究分析了2008至2015年间319例患者共438次脑深部电刺激脉冲发生器更换手术的并发症。总体并发症率为8.90%,包括感染2.74%、局部伤口糜烂1.37%、硬件故障2.51%、局部出血0.68%等,1例80岁患者术后4天死于心力衰竭加重。作者认为患者和医生应了解该并发症率以助选择可充电脉冲发生器。
要点
- 脉冲发生器更换手术总体并发症率为8.90%。
- 感染发生率为2.74%,硬件故障为2.51%。
- 1例80岁患者术后4天死于心力衰竭加重。
核心结论
- 脉冲发生器更换手术虽常被视为小手术,但在本中心并发症率约为9%。
- 患者和医生应了解脉冲发生器更换手术的并发症率,该信息可能有助于选择可充电脉冲发生器。
研究设计与样本
- 单中心回顾性研究,回顾性分析2008至2015年本科室所有接受脉冲发生器更换手术患者的医疗记录。
- 共319例患者,更换438个脉冲发生器,最短随访时间为19个月。
主要发现
- 总体并发症率为8.90%。
- 感染发生于12例患者(2.74%),局部伤口糜烂6例(1.37%),硬件故障11例(2.51%),局部出血3例(0.68%)。
- 脉冲发生器重复固定2例(0.46%),连接电缆牵拉需手术翻修2例(0.46%),因初始位置导致患者不适而将脉冲发生器置于腹部或更换为更小装置2例(0.46%)。
- 1例80岁患者(0.23%)心力衰竭严重加重,于脉冲发生器更换术后4天死亡。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Nonrechargeable deep brain stimulation impulse generators (IGs) with low or empty battery status require surgical IG exchange several years after initial implantation. The aim of this study was to investigate complication rates after IG exchange surgery and identify risk factors. METHODS: We retrospectively analyzed complications following IG exchange surgery from 2008 to 2015 in our department. Medical records of all patients who underwent IG exchange surgery were systematically reviewed. The shortest follow-up time was 19 months. RESULTS: From 2008 to 2015, 438 IGs were exchanged in 319 patients. Overall complication rate was 8.90%. Infection developed in 12 patients (2.74%). Six patients (1.37%) experienced local wound erosions. Hardware malfunctions were present in 11 patients (2.51%), and local hemorrhage was observed in 3 cases (0.68%). Repeated fixation of the IG was required in 2 patients (0.46%). Traction of the connecting cables necessitated surgical revision in 2 patients (0.46%). In 2 cases (0.46%), the IG was placed abdominally or exchanged for a smaller device owing to patient discomfort resulting from the initial positioning. One 80-year-old patient (0.23%) had severely worsening heart failure and died 4 days after IG exchange surgery. CONCLUSIONS: IG exchange surgery, although often considered a minor surgery, was associated with a complication rate of approximately 9% in our center. Patients and physicians should understand the complication rates associated with IG exchange surgery because this information might facilitate selection of a rechargeable IG.
引用
- Helmers AK, Lübbing I, Birkenfeld F, et al. Complications of Impulse Generator Exchange Surgery for Deep Brain Stimulation: A Single-Center, Retrospective Study. World Neurosurg. 2018. PMID: 29454121. doi:10.1016/j.wneu.2018.01.183
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。