扫描光纤内窥镜提高对浸润性胶质瘤边界处5-氨基乙酰丙酸诱导原卟啉IX荧光的检测
该研究在体外GL261-Luc2细胞及小鼠脑内浸润性胶质瘤生长阶段模型中,评估扫描光纤内窥镜(SFE)检测5-氨基乙酰丙酸诱导的原卟啉IX(PpIX)荧光的能力,并与配备荧光模块的手术显微镜(OPMI)及台式共聚焦激光扫描显微镜(CLSM)比较。结果显示SFE在细胞水平与CLSM成像相关,对荧光脑区累积PpIX的敏感性高于OPMI(P < 0.01),且将因光漂白导致肿瘤-背景对比度下降前的成像时间延长超过6倍。
要点
- SFE在细胞水平与CLSM的PpIX成像结果相关。
- SFE对荧光脑区累积PpIX的敏感性高于OPMI(P < 0.01)。
- SFE将光漂白导致对比度下降前的成像时间延长超过6倍。
核心结论
- SFE提供了在细胞分辨率下观察PpIX荧光肿瘤区域的新内窥镜能力,可能有助于在当前检测技术无法可靠显像时对胶质瘤等肿瘤的5-氨基乙酰丙酸标记进行准确成像。
- SFE对低PpIX浓度的敏感性显著高于OPMI,这与识别恶性胶质瘤前缘或转移细胞或治疗低级别胶质瘤相关。
- 这一新应用有潜力改善手术结局。
研究设计与样本
- 在体外GL261-Luc2细胞及植入小鼠脑内、处于浸润性胶质瘤生长阶段(模拟残留肿瘤)的模型中评估5-氨基乙酰丙酸诱导的PpIX荧光。
- 比较SFE、OPMI、CLSM和组织病理学对正常脑与肿瘤区域高、低PpIX浓度的术中荧光。
主要发现
- SFE的PpIX成像在细胞水平与CLSM相关。
- PpIX在正常脑细胞中积聚,但显著少于胶质瘤细胞。
- SFE对荧光脑区累积PpIX的敏感性高于OPMI(P < 0.01)。
- SFE将肿瘤-背景对比度因光漂白而下降前的成像时间延长超过6倍(>6×)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Fluorescence-guided surgery with protoporphyrin IX (PpIX) as a photodiagnostic marker is gaining acceptance for resection of malignant gliomas. Current wide-field imaging technologies do not have sufficient sensitivity to detect low PpIX concentrations. We evaluated a scanning fiber endoscope (SFE) for detection of PpIX fluorescence in gliomas and compared it to an operating microscope (OPMI) equipped with a fluorescence module and to a benchtop confocal laser scanning microscope (CLSM). METHODS: 5-Aminolevulinic acid-induced PpIX fluorescence was assessed in GL261-Luc2 cells in vitro and in vivo after implantation in mouse brains, at an invading glioma growth stage, simulating residual tumor. Intraoperative fluorescence of high and low PpIX concentrations in normal brain and tumor regions with SFE, OPMI, CLSM, and histopathology were compared. RESULTS: SFE imaging of PpIX correlated to CLSM at the cellular level. PpIX accumulated in normal brain cells but significantly less than in glioma cells. SFE was more sensitive to accumulated PpIX in fluorescent brain areas than OPMI (P < 0.01) and dramatically increased imaging time (>6×) before tumor-to-background contrast was diminished because of photobleaching. CONCLUSIONS: SFE provides new endoscopic capabilities to view PpIX-fluorescing tumor regions at cellular resolution. SFE may allow accurate imaging of 5-aminolevulinic acid labeling of gliomas and other tumor types when current detection techniques have failed to provide reliable visualization. SFE was significantly more sensitive than OPMI to low PpIX concentrations, which is relevant to identifying the leading edge or metastasizing cells of malignant glioma or to treating low-grade gliomas. This new application has the potential to benefit surgical outcomes.
引用
- Belykh E, Miller EJ, Hu D, et al. Scanning Fiber Endoscope Improves Detection of 5-Aminolevulinic Acid-Induced Protoporphyrin IX Fluorescence at the Boundary of Infiltrative Glioma. World Neurosurg. 2018. PMID: 29408716. doi:10.1016/j.wneu.2018.01.151
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。