该回顾性分析评估了2012年1月至2016年6月间71例分流手术中腔内内镜辅助导管放置的可行性、术后影像及分流失败率;68/71例按计划完成,术后影像排除完全错位,但仅64.7%(44/68例)达到最佳位置,4例因功能障碍需翻修(失败率5.8%),另5例因感染并发症或伤口愈合障碍需拔除,未发现导管并发症与次优位置或脑室裂隙/变形之间的直接相关性。

要点

  • 71例手术中68例按计划完成腔内内镜操作,术后影像可排除所有导管完全错位。
  • 最佳导管位置仅见于64.7%(44/68例),超过三分之一病例存在次优放置。
  • 4例导管因功能障碍需翻修(失败率5.8%),未观察到次优位置对后续分流失败的负面影响。

核心结论

  • 腔内内镜的多功能性反映了该技术的优势,无论何种手术均可避免意外放置或完全错位。
  • 然而超过三分之一病例出现次优导管放置,但未观察到对后续分流失败的负面影响。

研究设计与样本

  • 回顾性分析,2012年1月至2016年6月间71例手术中应用腔内内镜。
  • 应用场景包括首次分流或脑脊液储液囊置入(n=38)、近端分流失败翻修(n=13)及各种脑室内支架置入(n=20)。
  • 术后影像采用四点量表对导管位置进行分级,所有患者定期随访(平均31.6个月)以识别分流失败。

主要发现

  • 腔内内镜在68/71例手术中按计划完成。
  • 术后影像可排除所有导管完全错位,但最佳位置仅见于64.7%(44/68例)。
  • 4例导管因功能障碍需翻修(失败率5.8%),另5例因感染并发症或伤口愈合障碍需拔除。
  • 未观察到导管并发症与次优导管位置之间的直接相关性,脑室裂隙或变形也未证明是所观察并发症的危险因素。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The long-term function of a cerebral shunt is directly influenced by the placement of the ventricle catheter. In this work, an intra-luminal endoscope for best possible catheter positioning was used. Practicability, postoperative imaging, and shunt failure rates were retrospectively evaluated. METHODS: Between January 2012 and June 2016, an intra-catheter endoscope was applied in 71 procedures. Endoscopic technique was used for catheter placement in first-time shunting or cerebrospinal fluid reservoir insertion (n = 38), revision surgery in proximal shunt failure (n = 13), and various intraventricular stenting procedures (n = 20). Catheter positioning was graded on postoperative imaging using a four-point scale. All patients were regularly followed up (mean, 31.6 months) to recognize shunt failures. RESULTS: Endoscopic application could be completed as intended in 68 of 71 procedures. Postoperative imaging could exclude complete misplacement of all catheters, but optimal positioning was only achieved in 64.7% (44/68 cases). Four catheters had to be revised due to malfunction (failure rate, 5.8%). Another five catheters had to be removed due to infectious complications or wound-healing disorders. Direct correlations between catheter complications and suboptimal catheter positioning were not seen. Slit or distorted ventricles also did not prove to be a risk factor for the observed complications. CONCLUSIONS: Versatile application possibilities of the intra-catheter endoscope reflect the advantages of the technique. Independent of the performed procedure, unintended positionings or even complete catheter misplacements could be avoided. However, in more than one-third of all cases, suboptimal catheter placements became obvious. Interestingly, negative influences on later shunt failures were not seen.

来源:Acta Neurochir (Wien),查看原文。版权归原作者所有。

引用

  1. Antes S, Linsler S, Salah M, et al. Intra-catheter endoscopy for various shunting procedures-a retrospective analysis on surgical practicability, catheter placement, and failure rates. Acta Neurochir (Wien). 2017. PMID: 28695446. doi:10.1007/s00701-017-3264-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。