使用患者特异性三维打印复制品优化四维相位对比磁共振成像协议用于血流导向装置置入前后的在体成像

Four-Dimensional Phase Contrast Magnetic Resonance Imaging Protocol Optimization Using Patient-Specific 3-Dimensional Printed Replicas for In Vivo Imaging Before and After Flow Diverter Placement.
PubMed 文献 · World Neurosurg · 2017-06-15

研究用三维打印的2例脑动脉瘤复制品连接磁共振兼容连续流环路,优化四维相位对比磁共振成像(4D pcMRI)协议使成像时间不超过10分钟。6例动脉瘤(4.7–13.8 mm)患者用优化协议成像,其中3例置入Pipeline栓塞装置(PED)后复查。PED置入后瘤内有序血流被破坏,瘤内流速平均降低24.5%(范围12.9–31.5%)。

要点

  • 三维打印动脉瘤复制品可用于优化4D pcMRI采集参数,成像时间不超过10分钟。
  • 6例脑动脉瘤患者用优化协议成像,主要血流模式均显示良好。
  • PED置入后3例患者瘤内流速平均降低24.5%(范围12.9–31.5%)。

核心结论

  • 使用优化采集参数,脑动脉瘤4D pcMRI血流成像可在适合临床成像应用的时间范围内完成,从而能够量化血流导向装置治疗后的瘤内血流变化。

研究设计与样本

  • 用熔融沉积快速成型(三维打印)制作2例脑动脉瘤的三维复制品,连接磁共振兼容连续流环路行4D pcMRI成像,优化采集参数使成像时间不超过10分钟。
  • 6例脑动脉瘤患者(动脉瘤大小4.7–13.8 mm)用优化4D pcMRI协议成像,其中3例在Pipeline栓塞装置(PED)治疗后重复4D pcMRI检查。

主要发现

  • 所有病例主要血流模式均显示良好;较小动脉瘤因空间分辨率有限而具挑战性,较大动脉瘤因存在低流速区域导致流速敏感图像对比有限。
  • PED置入后瘤内有序血流被破坏,瘤内流速平均降低24.5%(范围12.9–31.5%)。
  • 探索性统计分析显示,流入速度变化与治疗后瘤内流速之间呈正向显著相关(P < 0.01)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Hemodynamics in cerebral aneurysms are currently investigated toward clinical efficacy using nonstandardized computational simulation techniques. At the same time, flow patterns and velocities are accessible by 4-dimensional phase contrast magnetic resonance imaging (4D pcMRI). Complexity of protocol design and imaging duration has limited the use of this technique in clinical imaging. A new approach is presented to overcome these limitations. METHODS: Three-dimensional (3D) replicas of 2 cerebral aneurysms were fabricated by fused deposition prototyping (3D printing) and imaged using 4D pcMRI while connected to a magnetic resonance imaging-compatible continuous flow loop. Acquisition parameters were optimized with imaging times not to exceed 10 minutes. Six patients harboring cerebral aneurysms with sizes ranging from 4.7 to 13.8 mm were imaged with the optimized 4D pcMRI protocol. After treatment with the pipeline embolization device (PED), 4D pcMRI examinations were repeated in 3 patients. RESULTS: In all cases, major flow patterns were visualized well; smaller aneurysms posed a challenge because of limited spatial resolution, whereas larger aneurysms contained regions of low velocity resulting in limited contrast in the flow-sensitive images. After PED placement, ordered aneurysmal flow was disrupted and intra-aneurysmal velocity was reduced on average by 24.5% (range, 12.9-31.5%). Exploratory statistical analysis yielded a positive significant correlation (P < 0.01) between changes in inflow velocity and posttreatment intra-aneurysmal flow velocity. CONCLUSIONS: 4D pcMRI flow imaging in cerebral aneurysms within a time frame suitable for clinical imaging applications is feasible with optimized acquisition parameters, thereby enabling quantification of intra-aneurysmal flow changes after flow diverter device treatment.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Karmonik C, Anderson JR, Elias S, et al. Four-Dimensional Phase Contrast Magnetic Resonance Imaging Protocol Optimization Using Patient-Specific 3-Dimensional Printed Replicas for In Vivo Imaging Before and After Flow Diverter Placement. World Neurosurg. 2017. PMID: 28624566. doi:10.1016/j.wneu.2017.06.042