全植入式双向神经接口在5例帕金森病患者中的慢性多部位脑记录经验
在医生发起的美国食品药品监督管理局研究性器械豁免下,5例帕金森病患者植入Activa PC+S系统,于运动皮层硬膜下和丘脑底核放置四极电极,进行为期1年的多时间点慢性多部位场电位记录。结果显示无严重手术并发症或脑深部电刺激治疗中断,皮层和丘脑底核信号随时间相对稳定,尽管电极阻抗逐渐增加;大多数但非全部记录中可识别运动皮层特定频段的典型运动相关变化。结论认为人类慢性多部位场电位采集可行。
要点
- 5例帕金森病患者植入全植入式双向神经接口,实现慢性多部位场电位记录。
- 无严重手术并发症或脑深部电刺激治疗中断。
- 皮层和丘脑底核信号随时间相对稳定,电极阻抗逐渐增加。
核心结论
- 人类慢性多部位场电位采集是可行的。
- 该研究工具为将脑网络动力学发现转化为改进的神经刺激范式提供了平台。
研究设计与样本
- 在医生发起的美国食品药品监督管理局研究性器械豁免下,5例帕金森病患者植入Activa PC+S系统(Medtronic Inc.)。
- 装置连接置于运动皮层硬膜下空间的四极电极用于皮层脑电位记录,并连接置于丘脑底核的四极电极用于治疗性刺激和局部场电位记录。
- 在1年期间对每例患者的大脑进行多个时间点的记录。
主要发现
- 无严重手术并发症或脑深部电刺激治疗中断。
- 皮层和丘脑底核信号随时间相对稳定,尽管电极阻抗逐渐增加。
- 大多数但非全部记录中可识别运动皮层特定频段的典型运动相关变化。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE Dysfunction of distributed neural networks underlies many brain disorders. The development of neuromodulation therapies depends on a better understanding of these networks. Invasive human brain recordings have a favorable temporal and spatial resolution for the analysis of network phenomena but have generally been limited to acute intraoperative recording or short-term recording through temporarily externalized leads. Here, the authors describe their initial experience with an investigational, first-generation, totally implantable, bidirectional neural interface that allows both continuous therapeutic stimulation and recording of field potentials at multiple sites in a neural network. METHODS Under a physician-sponsored US Food and Drug Administration investigational device exemption, 5 patients with Parkinson's disease were implanted with the Activa PC+S system (Medtronic Inc.). The device was attached to a quadripolar lead placed in the subdural space over motor cortex, for electrocorticography potential recordings, and to a quadripolar lead in the subthalamic nucleus (STN), for both therapeutic stimulation and recording of local field potentials. Recordings from the brain of each patient were performed at multiple time points over a 1-year period. RESULTS There were no serious surgical complications or interruptions in deep brain stimulation therapy. Signals in both the cortex and the STN were relatively stable over time, despite a gradual increase in electrode impedance. Canonical movement-related changes in specific frequency bands in the motor cortex were identified in most but not all recordings. CONCLUSIONS The acquisition of chronic multisite field potentials in humans is feasible. The device performance characteristics described here may inform the design of the next generation of totally implantable neural interfaces. This research tool provides a platform for translating discoveries in brain network dynamics to improved neurostimulation paradigms. Clinical trial registration no.: NCT01934296 (clinicaltrials.gov).
引用
- Swann NC, de Hemptinne C, Miocinovic S, et al. Chronic multisite brain recordings from a totally implantable bidirectional neural interface: experience in 5 patients with Parkinson's disease. J Neurosurg. 2017. PMID: 28409730. doi:10.3171/2016.11.JNS161162
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。