基于MRI的图像引导用于脊柱间质激光热疗:技术要点与准确性
本研究将术中脊柱MRI配准至神经导航系统,用于引导激光间质热疗探针置入治疗转移性脊髓压迫。13例患者共置入47根激光探针,导航预测位置与实际接入套管位置的中位绝对偏差为0.7 mm(范围0–3.2 mm),术中无损伤或不良事件发生。作者认为该方法可行,能清晰显示脊髓,提高硬膜外腔激光消融的安全性与工作流程。
要点
- 将术中脊柱MRI配准至导航系统,可清晰显示脊髓等软组织。
- 13例患者置入47根激光探针,位置偏差中位0.7 mm(范围0–3.2 mm)。
- 术中无损伤或不良事件发生,作者认为该方法可行。
核心结论
- 本研究证明基于MRI的图像引导用于脊柱转移瘤激光间质热疗是可行的。
- 作者的方法能清晰显示脊髓,提高硬膜外腔激光消融的安全性与工作流程。
- 结果可外推至其他适应证,包括活检或脊髓附近积液引流。
研究设计与样本
- 患者在全麻下俯卧于MRI检查台,背部皮肤贴基准标记,并用塑料托架支撑MRI线圈。
- 将感兴趣区的T2加权MRI序列导出至标准导航系统,皮肤缝合参考阵列后对基准标记进行表面匹配。
- 导航下Jamshidi针推进至接触后部结构,术中透视确认位置后进入硬膜外腔目标,保存最终位置截图。
- 将Jamshidi针更换为MRI兼容接入套管,获取每根接入套管所在精确轴位平面的MRI,并与定位时保存的截图比较。
- 测量轨迹间以毫米计的偏差以评估图像引导的准确性;共13例患者置入47根激光探针。
主要发现
- 导航系统预测位置与接入套管实际位置之间偏差的中位绝对值为0.7 mm(范围0–3.2 mm)。
- 操作过程中无损伤或不良事件发生。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE Image guidance for spinal procedures is based on 3D-fluoroscopy or CT, which provide poor visualization of soft tissues, including the spinal cord. To overcome this limitation, the authors developed a method to register intraoperative MRI (iMRI) of the spine into a neuronavigation system, allowing excellent visualization of the spinal cord. This novel technique improved the accuracy in the deployment of laser interstitial thermal therapy probes for the treatment of metastatic spinal cord compression. METHODS Patients were positioned prone on the MRI table under general anesthesia. Fiducial markers were applied on the skin of the back, and a plastic cradle was used to support the MRI coil. T2-weighted MRI sequences of the region of interest were exported to a standard navigation system. A reference array was sutured to the skin, and surface matching of the fiducial markers was performed. A navigated Jamshidi needle was advanced until contact was made with the dorsal elements; its position was confirmed with intraoperative fluoroscopy prior to advancement into a target in the epidural space. A screenshot of its final position was saved, and then the Jamshidi needle was exchanged for an MRI-compatible access cannula. MRI of the exact axial plane of each access cannula was obtained and compared with the corresponding screenshot saved during positioning. The discrepancy in millimeters between the trajectories was measured to evaluate accuracy of the image guidance RESULTS Thirteen individuals underwent implantation of 47 laser probes. The median absolute value of the discrepancy between the location predicted by the navigation system and the actual position of the access cannulas was 0.7 mm (range 0-3.2 mm). No injury or adverse event occurred during the procedures. CONCLUSIONS This study demonstrates the feasibility of image guidance based on MRI to perform laser interstitial thermotherapy of spinal metastasis. The authors' method permits excellent visualization of the spinal cord, improving safety and workflow during laser ablations in the epidural space. The results can be extrapolated to other indications, including biopsies or drainage of fluid collections near the spinal cord.
引用
- Tatsui CE, Nascimento CNG, Suki D, et al. Image guidance based on MRI for spinal interstitial laser thermotherapy: technical aspects and accuracy. J Neurosurg Spine. 2017. PMID: 28186470. doi:10.3171/2016.9.SPINE16475
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。