使用加速度传感器在脑深部电刺激手术中进行僵直评估的新型辅助方法
本研究在3例患者术后随访及9例帕金森病双侧脑深部电刺激术中应用加速度传感器评估僵直变化。结果显示加速度评估可重复,188个刺激位点中可找到抑制僵直电流振幅的位点数从主观评估的144个增至170个,加速度评估所需振幅更低(平均0.9±0.6 mA对1.4±0.6 mA,p<0.001),14个植入位置选择中仅2个与主观评估相同。
要点
- 加速度传感器数据提取的指标对同一评估者具有可重复性。
- 188个刺激位点中,加速度评估使可找到抑制僵直振幅的位点数从144增至170。
- 加速度评估的僵直缓解振幅平均0.9±0.6 mA,低于主观评估的1.4±0.6 mA(p<0.001)。
核心结论
- 本文描述并验证了一种在脑深部电刺激术中刺激测试时使用加速度传感器评估僵直的新型辅助方法。
- 初步结果表明该方法可能成为优化脑深部电刺激导线放置的临床有用辅助手段,并可能成为寻找帕金森病最佳脑深部电刺激靶点的新工具。
研究设计与样本
- 方法可靠性在3例患者的术后随访检查中测试。
- 术中可行性在9例帕金森病患者的双侧脑深部电刺激手术中研究。
- 从加速度计数据中提取相关结局指标计算僵直变化,并与神经科医生识别的振幅进行统计学比较。
主要发现
- 加速度计数据提取的结局指标对同一评估者可重复,提供术中刺激测试时僵直变化的可靠评估。
- 分析的188个刺激位点中,考虑加速度评估时找到抑制僵直振幅的位点数从144增至170。
- 加速度评估的僵直缓解振幅平均0.9±0.6 mA,显著低于主观评估的平均1.4±0.6 mA(p<0.001)。
- 14个脑深部电刺激导线植入位置选择中,仅2个在加速度评估与主观评估间相同。
- 解剖位置比较显示,Forel区刺激缓解僵直的振幅与丘脑底核刺激相似,但副作用更少。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE Despite the widespread use of deep brain stimulation (DBS) for movement disorders such as Parkinson's disease (PD), the exact anatomical target responsible for the therapeutic effect is still a subject of research. Intraoperative stimulation tests by experts consist of performing passive movements of the patient's arm or wrist while the amplitude of the stimulation current is increased. At each position, the amplitude that best alleviates rigidity is identified. Intrarater and interrater variations due to the subjective and semiquantitative nature of such evaluations have been reported. The aim of the present study was to evaluate the use of an acceleration sensor attached to the evaluator's wrist to assess the change in rigidity, hypothesizing that such a change will alter the speed of the passive movements. Furthermore, the combined analysis of such quantitative results with anatomy would generate a more reproducible description of the most effective stimulation sites. METHODS To test the reliability of the method, it was applied during postoperative follow-up examinations of 3 patients. To study the feasibility of intraoperative use, it was used during 9 bilateral DBS operations in patients suffering from PD. Changes in rigidity were calculated by extracting relevant outcome measures from the accelerometer data. These values were used to identify rigidity-suppressing stimulation current amplitudes, which were statistically compared with the amplitudes identified by the neurologist. Positions for the chronic DBS lead implantation that would have been chosen based on the acceleration data were compared with clinical choices. The data were also analyzed with respect to the anatomical location of the stimulating electrode. RESULTS Outcome measures extracted from the accelerometer data were reproducible for the same evaluator, thus providing a reliable assessment of rigidity changes during intraoperative stimulation tests. Of the 188 stimulation sites analyzed, the number of sites where rigidity-suppressing amplitudes were found increased from 144 to 170 when the accelerometer evaluations were considered. In general, rigidity release could be observed at significantly lower amplitudes with accelerometer evaluation (mean 0.9 ± 0.6 mA) than with subjective evaluation (mean 1.4 ± 0.6 mA) (p < 0.001). Of 14 choices for the implant location of the DBS lead, only 2 were the same for acceleration-based and subjective evaluations. The comparison across anatomical locations showed that stimulation in the fields of Forel ameliorates rigidity at similar amplitudes as stimulation in the subthalamic nucleus, but with fewer side effects. CONCLUSIONS This article describes and validates a new assistive method for assessing rigidity with acceleration sensors during intraoperative stimulation tests in DBS procedures. The initial results indicate that the proposed method may be a clinically useful aid for optimal DBS lead placement as well as a new tool in the ongoing scientific search for the optimal DBS target for PD.
引用
- Shah A, Coste J, Lemaire JJ, et al. A novel assistive method for rigidity evaluation during deep brain stimulation surgery using acceleration sensors. J Neurosurg. 2016. PMID: 27982769. doi:10.3171/2016.8.JNS152770
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。