Flucropen:一种用于脑肿瘤立体定向针吸活检术中检测荧光素荧光的简易低成本装置

The Fluoropen: a simple low-cost device to detect intraoperative fluorescein fluorescence in stereotactic needle biopsy of brain tumors.
PubMed 文献 · Acta Neurochir (Wien) · 2016-12-10

本研究开发了Flucropen装置,用于荧光素荧光引导的立体定向针吸活检中检测脑肿瘤组织荧光。在连续4例患者的6份样本中,Flucropen与荧光显微镜的检测一致性为100%。该低成本独立装置可即时确认诊断样本,避免术中冰冻切片,但对非强化肿瘤及术前使用地塞米松者需谨慎。

要点

  • Flucropen与荧光显微镜检测荧光的一致性为100%。
  • 该装置可即时确认诊断样本,避免术中冰冻切片。
  • 非强化肿瘤及术前使用地塞米松者需谨慎使用荧光引导活检。

核心结论

  • Flucropen是一种低成本、简易的独立荧光检测装置,可即时确认诊断样本,避免术中冰冻切片,从而加快立体定向针吸活检。
  • 对于非强化肿瘤及术前使用地塞米松的患者,荧光素荧光引导的立体定向针吸活检需谨慎使用。

研究设计与样本

  • 概念验证研究,连续纳入4例接受荧光素荧光引导无框架立体定向脑肿瘤活检的患者,共获取6份样本,每份样本用Flucropen检测荧光并与具备荧光可视化能力的显微镜比较。

主要发现

  • 从肿瘤强化部分获取的5份样本中,4份(80%)在配备荧光模块的显微镜和Flucropen下均显示荧光。
  • 1份非强化病灶样本在配备荧光模块的显微镜和Flucropen下均无荧光。
  • Flucropen与配备荧光模块的显微镜显示100%一致性。

临床怎么用

  • 作者认为Flucropen可即时确认诊断样本,避免术中冰冻切片,从而加快立体定向针吸活检。
  • 作者指出,对于非强化肿瘤及术前使用地塞米松的患者,荧光素荧光引导的立体定向针吸活检需谨慎使用。

局限性

  • 概念验证研究仅纳入连续4例患者、共6份样本。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The use of fluorescein fluorescence-guided stereotactic needle biopsy has been shown to improve diagnostic accuracy and to expedite operative procedure in the stereotactic needle biopsy of high-grade gliomas. We developed a device (Fluoropen) for detecting fluorescence in brain tumor tissues obtained by fluorescein fluorescence-guided stereotactic needle biopsy. METHODS: The Fluoropen is a device consisting of a light source fitted with color filters to create the required emission and visualization wavelengths. The proof-of-concept study consisted of four consecutive patients who underwent fluorescein fluorescence-guided frameless stereotactic biopsy of brain tumor. Each sample was examined for the presence of fluorescence using the Fluoropen and compared with a microscope with fluorescence visualization capability. RESULTS: A total of six samples were obtained from four stereotactic needle biopsy procedures. Four out of five samples (80%) taken from the contrast-enhancing part of the tumors were shown to be fluorescent under the microscope fitted with fluorescence module and the Fluoropen. One non-contrast enhancing lesion was non-fluorescent using both the microscope fitted with fluorescence module and the Fluoropen. The Fluoropen was shown to have 100% concordance with the microscope fitted with fluorescence module. CONCLUSIONS: The Fluoropen is a low-cost and simple standalone device for the detection of fluorescein fluorescence that can expedite stereotactic needle biopsy by providing instant confirmation of the diagnostic sample and therefore avoid the need for an intraoperative frozen section. In patients with non-contrast enhancing tumors and those who were pre-treated with dexamethasone prior to surgery, fluorescein fluorescence-guided stereotactic needle biopsy will need to be used with caution.

来源:Acta Neurochir (Wien),查看原文。版权归原作者所有。

引用

  1. Thien A, Rao JP, Ng WH, et al. The Fluoropen: a simple low-cost device to detect intraoperative fluorescein fluorescence in stereotactic needle biopsy of brain tumors. Acta Neurochir (Wien). 2016. PMID: 27943078. doi:10.1007/s00701-016-3041-8