三维评估高密度栓塞材料对动静脉畸形伽玛刀放射外科靶区吸收剂量的影响

Three-dimensional assessment of the effects of high-density embolization material on the absorbed dose in the target for Gamma Knife radiosurgery of arteriovenous malformations.
PubMed 文献 · J Neurosurg · 2016-01-01

本研究通过三维打印体模填充组织等效聚合物凝胶并注入Onyx 18模拟动静脉畸形栓塞,比较TMR 10与卷积算法计算的伽玛刀剂量分布,并分析2例既往接受伽玛刀治疗患者的剂量差异。体模分析显示靶区平均剂量差异不大于6%,患者病例差异为5%,栓塞材料附近局部剂量差异较大,但总体剂量降低不超过5%。

要点

  • 体模中靶区平均剂量差异(TMR 10与卷积算法)不大于6%,患者病例为5%。
  • 剂量差异较大区域主要位于栓塞材料处或其邻近。
  • 体模与患者研究均显示栓塞材料所致剂量降低不超过5%,对闭塞率的临床影响预计较小。

核心结论

  • 体模与患者研究均显示,靶区附近栓塞材料导致的剂量降低不超过5%,对闭塞率的临床影响预计较小。

研究设计与样本

  • 体模研究:3D打印圆柱体模(直径16 cm、长12 cm,带直径16 cm半球顶)填充组织等效聚合物凝胶,注入Onyx 18至悬浮的AVM模型内,以Leksell框架固定后行CT扫描(120 kV、1.5 mm层厚),用GammaPlan 10.1.1制定计划(11个8 mm等中心,处方剂量20 Gy至50%等剂量线),分别用TMR 10和卷积算法计算三维剂量分布并比较。
  • 患者研究:对2例既往接受伽玛刀放射外科治疗的患者应用相同分析方法评估剂量效应。

主要发现

  • 体模分析显示AVM靶区TMR 10与卷积算法之间的平均剂量差异不大于6%。
  • 伽玛刀病例的剂量差异为5%。
  • 等剂量线图上观察到小区域存在较大剂量差异,这些差异大多位于栓塞材料处或其邻近。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE Arteriovenous malformation (AVM) is an intracranial vascular disorder. Gamma Knife radiosurgery (GKRS) is used in conjunction with intraarterial embolization to eradicate the nidus of AVMs. Clinical results indicate that patients with prior embolization tend to gain less benefit from GKRS. The authors hypothesized that this was partly caused by dosimetric deficiency. The actual dose delivered to the target may be smaller than the intended dose because of increased photon attenuation by high-density embolic materials. The authors performed a phantom-based study to quantitatively evaluate the 3D dosimetric effect of embolic material on GKRS. METHODS A 16-cm-diameter and 12-cm-long cylindrical phantom with a 16-cm-diameter hemispherical dome was printed by a 3D printer. The phantom was filled with radiologically tissue-equivalent polymer gel. To simulate AVM treatment with embolization, phantoms contained Onyx 18. The material was injected into an AVM model, which was suspended in the polymer gel. The phantom was attached to a Leksell frame by standard GK fixation method, using aluminum screws, for imaging. The phantom was scanned by a Phillips CT scanner with the standard axial-scanning protocol (120 kV and 1.5-mm slice thickness). CT-based treatment planning was performed with the GammaPlan treatment planning system (version 10.1.1). The plan was created to cover a fictitious AVM target volume near the embolization areas with eleven 8-mm shots and a prescription dose of 20 Gy to 50% isodose level. Dose distributions were computed using both tissue maximum ratio (TMR) 10 and convolution dose-calculation algorithms. These two 3D dose distributions were compared using an in-house program. Additionally, the same analysis method was applied to evaluate the dosimetric effects for 2 patients previously treated by GKRS. RESULTS The phantom-based analyses showed that the mean dose difference between TMR 10 and convolution doses of the AVM target was no larger than 6%. The difference for GKRS cases was 5%. There were small areas where a large dose difference was observed on the isodose line plots, and those differences were mostly at or in the vicinity of the embolization materials. CONCLUSIONS The results of both the phantom and patient studies showed a dose reduction no larger than 5% due to the embolization material placed near the target. Although the comparison of 3D dose distributions indicated small local effects of the embolic material, the clinical impact on the obliteration rate is expected to be small.