文献速递小儿神经外科与脑积水

对比增强分流管系列(分流管造影)在检测分流管梗阻方面优于其他分流管影像学方法

Contrast-enhanced shunt series ("shuntography") compare favorably to other shunt imaging modalities in detecting shunt occlusion.
PubMed 文献 · Acta Neurochir (Wien) · 2016-11-05

本研究回顾性再评估107例对比增强分流管系列检查并复习患者病史,计算其预测分流管翻修手术的效用参数。41例检查阳性,其中30天内35例翻修成功、3例翻修后症状未缓解、2例分流管通畅;66例阴性检查中2例翻修、10例未尝试恢复通畅。对比增强分流管系列对分流管失效的灵敏度94.7%、特异度92.8%、阳性预测值87.8%、阴性预测值97.0%。作者结论认为该方法高度特异,阴性预测值超过头颅CT和平片分流管系列。

要点

  • 对比增强分流管系列对分流管失效的阴性预测值为97.0%,高于头颅CT和平片分流管系列。
  • 107例检查中41例阳性,梗阻主要位于脑室端(36.5%)和腹腔端(48.8%)。
  • 与放射性核素标记研究相比,对比增强分流管系列时空分辨率更好,可指导局部手术修复。

核心结论

  • 对比增强分流管系列是一种高度特异的检查,其阴性预测值超过头颅计算机断层扫描和平片分流管系列。
  • 与放射性核素标记研究相比,对比增强分流管系列时空分辨率更好,能够实现针对性的局部手术修复。

研究设计与样本

  • 回顾性再评估107例对比增强分流管系列检查并复习患者病史,定义结局参数以计算异常对比增强分流管系列预测翻修手术的效用。

主要发现

  • 107例对比增强分流管系列中41例检查阳性,梗阻主要位于脑室端导管(36.5%)和腹腔端导管(48.8%)。
  • 30天内完成35例成功翻修手术和3例症状未缓解的翻修手术,2例发现分流管通畅。
  • 66例阴性检查导致2例翻修手术,另有10例手术未尝试恢复通畅。
  • 30天后对比增强分流管系列识别分流管失效的特异度为92.8%,灵敏度为94.7%,阳性预测值为87.8%,阴性预测值为97.0%。

关键数值与推荐

  • 对比增强分流管系列识别分流管失效:特异度92.8%、灵敏度94.7%、阳性预测值87.8%、阴性预测值97.0%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Obstruction is a common cause of ventriculo-peritoneal shunt failure. Head computed tomography and plain x-ray examinations of shunt tubing ("shunt series") are routinely used in patients readmitted for reemerging symptoms but are of limited value. The validity of shunt series can be improved by applying contrast agent into the system (contrast-enhanced shunt series, a.k.a. a "shuntogram" or "shuntography"). We hypothesized that contrast-enhanced shunt series have a high predictive value for shunt revision surgeries. METHODS: We retrospectively re-evaluated 107 contrast-enhanced shunt series and reviewed the patient histories. We defined outcome parameters for calculating the utility of a pathological contrast-enhanced shunt series in predicting revision surgery. RESULTS: Of 107 contrast-enhanced shunt series, 41 examinations were positive for obstruction, mainly of the ventricular (36.5 %) and the peritoneal catheter (48.8 %). Within 30 days, 35 successful revision surgeries and 3 revision surgeries without resolution of symptoms were performed. In two cases the shunt tubing was found to be patent. Sixty-six negative examinations resulted in two revision surgeries, in addition to ten surgeries not attempting to restore patency. After 30 days, the specificity of contrast-enhanced shunt series for shunt failure identification was calculated at 92.8 %, the sensitivity at 94.7 %, the positive predictive value at 87.8 %, and the negative predictive value at 97.0 %. CONCLUSIONS: The contrast-enhanced shunt series method is a highly specific examination with a negative predictive value exceeding that of head computed tomography and plain shunt series. Compared to radionuclide marker studies, contrast-enhanced shunt series demonstrate better spatiotemporal resolution, enabling focused local surgical repair.

来源:Acta Neurochir (Wien),查看原文。版权归原作者所有。