文献速递功能神经外科

功能导向脑深部电刺激基于三维轨迹旋转几何推断多轨道靶点的立体定向精度与磁共振引导脑深部电刺激及结局的比较

Comparison of the Stereotactic Accuracies of Function-Guided Deep Brain Stimulation, Calculated Using Multitrack Target Locations Geometrically Inferred from Three-Dimensional Trajectory Rotations, and of Magnetic Resonance Imaging-Guided Deep Brain Stimulation and Outcomes.
PubMed 文献 · World Neurosurg · 2016-11-19

本研究回顾性评估172根脑深部电刺激电极的立体定向误差,比较功能导向(多轨道微电极记录+清醒刺激)与MRI引导两组。STN功能组误差1.4±0.7 mm,STN MRI组1.0±0.6 mm(P=0.06);功能组中误差较大的电极被选择性偏移轨道,86%(56/65)接近MRI靶点。GPi MRI组误差0.9±0.5 mm。误差较低与较好结局趋势相关(P=0.095)。

要点

  • 功能导向DBS的STN电极立体定向误差1.4±0.7 mm,大于MRI引导的1.0±0.6 mm(P=0.06)。
  • 功能导向组中误差较大的电极被选择性偏移轨道,86%(56/65)接近MRI靶点。
  • GPi MRI引导组立体定向误差为0.9±0.5 mm。

核心结论

  • 使用术中功能评估的多轨道DBS立体定向精度劣于MRI引导DBS,但功能导向DBS的轨道偏移可有效接近MRI靶点。

研究设计与样本

  • 回顾性评估2014年4月至2016年1月间用于DBS的172根电极的立体定向误差。
  • 靶点为丘脑底核(STN,139根电极)或苍白球内侧部(GPi,33根电极)。
  • 通过考虑三维立体定向环和弧旋转,几何计算偏移轨道靶点;立体定向误差用垂直于轨迹的欧氏距离计算。
  • 对24例随访1年的患者分析UPDRS第III部分运动结局及刺激改善百分比。

主要发现

  • 功能评估倾向于增加STN功能组(n=129)的立体定向误差至1.4±0.7 mm,高于STN MRI组(n=10)的1.0±0.6 mm(P=0.06)。
  • 立体定向误差较高的电极(n=65;1.6±0.7 mm)被选择性偏移轨道,其误差大于中心轨道电极(n=64;1.2±0.7 mm)(P<0.001)。
  • 偏移轨道的电极中86%(56/65根)接近MRI靶点。
  • 较低的立体定向误差与较好结局趋势相关(P=0.095);与MRI靶点距离>2.5 mm与较差结局趋势相关(P=0.087)。
  • GPi MRI组(n=33)的立体定向误差较低,为0.9±0.5 mm。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: In previous studies, multitrack trajectories in deep brain stimulation (DBS) have usually been approximated. Using a geometrically more accurate method, we compared the stereotactic accuracy of DBS with multitrack microelectrode recording and awake stimulation (function group) and MRI-guided DBS (MRI group). METHODS: One hundred and seventy-two leads used in DBS between April 2014 and January 2016 were evaluated for stereotactic errors. Targets were the subthalamic nucleus (STN, 139 leads) or globus pallidus interna (GPi, 33 leads). We geometrically calculated shifted-track targets by considering the three-dimensional stereotactic ring and arc rotations. Stereotactic errors were calculated using Euclidean distances perpendicular to trajectories. Motor outcomes according to the Unified Parkinson's Disease Rating Scale (UPDRS) part III, improvement percentages by stimulations, were analyzed in 24 patients with 1 year follow-ups. RESULTS: Functional evaluation tended to increase stereotactic errors in the STN function group (n = 129; 1.4 ± 0.7 mm) more than in the STN MRI group (n = 10; 1.0 ± 0.6 mm; P = 0.06). Leads with higher stereotactic errors (n = 65; 1.6 ± 0.7 mm; P < 0.001) than the center-track leads (n = 64; 1.2 ± 0.7 mm) were selectively track shifted. Track-shifted leads approached MRI targets in 86% (56/65 leads). Lower stereotactic errors tended to correlate with a better outcome (P = 0.095). Distances to MRI targets >2.5 mm tended to relate to a worse outcome (P = 0.087). Stereotactic errors were lower (n = 33; 0.9 ± 0.5 mm) in the GPi MRI group. CONCLUSIONS: Multitrack DBS using intraoperative functional evaluation resulted in worse stereotactic accuracy than did MRI-guided DBS. However, track shifts in function-guided DBS can approach MRI targets effectively.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Park SC, Lee CS, Kim SM, et al. Comparison of the Stereotactic Accuracies of Function-Guided Deep Brain Stimulation, Calculated Using Multitrack Target Locations Geometrically Inferred from Three-Dimensional Trajectory Rotations, and of Magnetic Resonance Imaging-Guided Deep Brain Stimulation and Outcomes. World Neurosurg. 2016. PMID: 27876666. doi:10.1016/j.wneu.2016.11.046

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。