文献速递小儿神经外科与脑积水

定量MRI评估脑积水患者基线分流流量:辅助临床决策的无创工具

Quantitative MRI to assess baseline shunt flow in patients with hydrocephalus: a noninvasive tool to aid clinical decision-making.
PubMed 文献 · J Neurosurg · 2026-10-16

该前瞻性观察队列研究纳入207例脑室分流患者,在常规MRI中加入相位对比(PC)序列测量脑脊液分流流量。平均流量为9.9±8 mL/小时,流量与年龄、性别、病因、阀门类型等因素无关;重复测量无显著差异(p=0.83)。PC-MRI识别出3例分流故障并避免1例不必要手术。

要点

  • PC-MRI可无创定量测量脑室分流流量,平均9.9±8 mL/小时。
  • 流量测量不受年龄、性别、病因、阀门类型等因素显著影响。
  • PC-MRI有助于识别分流故障,避免不必要手术。

核心结论

  • 在常规影像中加入PC-MRI可准确无创测量脑室分流流量,为分流功能评估提供补充信息并作为未来比较的基线。

研究设计与样本

  • 前瞻性观察队列研究,单中心,2023年1月至2024年1月,纳入207例脑室分流患者(107例男性,平均年龄11.3岁,范围8天至27岁),98%为VP分流。

主要发现

  • 平均脑脊液流量为9.9±8 mL/小时(范围0.5-54 mL/小时)。
  • 流量与年龄、性别、脑积水病因、阀门类型、分流置入时间间隔、BMI、额枕角比、额角或房部宽度均无显著差异。
  • 11例重复PC-MRI测量,初次与随访流量无显著差异(p=0.83)。
  • PC-MRI识别出3例分流故障,并避免1例不必要手术。

临床怎么用

  • 对脑室分流患者,可在常规MRI中加入PC-MRI定量测量分流流量,作为基线并辅助判断分流功能。

局限性

  • 单中心研究,重复测量样本量小(11例),需进一步验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Hydrocephalus is a common yet serious condition, and ventricular shunts are used for lifesaving CSF diversion. Existing methods for shunt evaluation provide correlative/qualitative information, leading to commonplace clinical diagnostic challenges. The aim of this study was to determine accurate and noninvasive measures of volumetric CSF flow rates through ventricular shunts using the phase-contrast (PC) sequence in routine MRI for pediatric patients with hydrocephalus. METHODS: The authors conducted a prospective observational cohort study of patients with ventricular shunts who underwent a clinically indicated brain MRI and the adjunct PC-MRI shunt flow sequence at a single center between January 2023 and January 2024. At the study institution, PC-MRI has been integrated as a supplementary imaging technique to quantify shunt flow alongside fast protocol MRI for shunted patients. Clinical and imaging data were collected. Data distributions and paired changes in flow measurements over time were assessed using appropriate statistical tests. RESULTS: Overall, 207 patients (107 male, mean age 11.3 years [range 8 days-27 years]) who underwent PC-MRI were included; 98% had a VP shunt. The mean CSF flow rate was 9.9 ± 8 mL/hour (range 0.5-54 mL/hour). No significant differences were noted by age, sex, etiology of hydrocephalus, valve type, time interval since shunt placement, BMI, fronto-occipital horn ratio, and frontal horn or atrial width. Repeat PC-MRI was performed in 11 patients, and there was no significant difference in flow measurements between initial and follow-up images (p = 0.83) among these patients. PC-MRI identified alterations that accurately demonstrated failure in 3 patients and avoided unnecessary surgery in a fourth patient. CONCLUSIONS: Addition of PC-MRI to routine imaging examinations in patients with ventricular shunts accurately and noninvasively measured CSF shunt flow. Flow rate data can provide supplementary information regarding shunt function and serve as a baseline for future comparison assessments. These findings suggest a role for routine acquisition of patient-specific baseline shunt flow measurement in shunted pediatric patients with hydrocephalus.

来源:J Neurosurg,查看原文。版权归原作者所有。

引用

  1. Ha JH, Gutierrez ME, Toliao JM, et al. Quantitative MRI to assess baseline shunt flow in patients with hydrocephalus: a noninvasive tool to aid clinical decision-making. J Neurosurg. 2026. PMID: 42887022. doi:10.3171/2026.5.JNS253071

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。