单极脑深部电刺激下诱导电流对功能磁共振成像可解释性的影响
该研究在标准化体模中测量两种电流控制型DBS系统在3-T MRI采集期间的输出稳定性,发现单极刺激下诱导电流波动达-3.2至+3.9 mA且频繁极性反转,而双极刺激输出稳定;作者据此推断单极DBS期间fMRI的BOLD反应可能反映失真而非程控刺激。
要点
- 单极DBS在3-T MRI期间出现显著电流波动和极性反转。
- 双极刺激在MRI期间输出稳定,偏差极小。
- 解读DBS期间fMRI数据时应优先采用双极配置。
核心结论
- MRI诱导电流显著改变单极DBS的有效输出,给同步fMRI采集引入不确定性。
- 双极配置可避免该混杂因素,在解读DBS期间fMRI数据时应优先采用。
研究设计与样本
- 在标准化体模中测量两种电流控制型DBS系统于3-T MRI采集期间的电输出。
- 刺激采用单极和双极配置,DBS系统处于开和关状态。
- 使用定制MRI兼容装置记录诱导电流,并在包括梯度密集fMRI序列的多种序列中量化峰值幅度。
- 所有数据归一化至基线输出,并用Cohen's d评估MRI诱导电流偏差的幅度。
主要发现
- 单极刺激在MRI期间出现显著电流波动,诱导幅度范围为-3.2至+3.9 mA,并频繁出现极性反转。
- 这些失真具有序列依赖性,在fMRI采集期间最为明显。
- 双极刺激输出稳定,与程控参数的偏差极小。
- 单极输出的变异性不能归因于阻抗变化,且在两种受试DBS系统中表现一致。
关键数值与推荐
- 单极刺激MRI期间诱导电流幅度范围为-3.2至+3.9 mA。
临床怎么用
- 解读单极DBS期间采集的fMRI数据时需考虑诱导电流导致的输出失真。
- 在DBS期间进行fMRI采集时优先选择双极配置以减少混杂。
局限性
- 研究未在人类受试者中直接进行功能成像。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The objective was to evaluate the stability of stimulation current delivered by deep brain stimulation (DBS) systems during MRI scanning and to assess whether configuration-dependent variability in induced current may undermine the interpretability of functional MRI (fMRI) acquired during active stimulation. METHODS: The authors measured the electrical output of 2 current-controlled DBS systems in a standardized phantom during 3-T MRI acquisition. Stimulation was delivered in both monopolar and bipolar configurations, with the DBS systems on and off. Induced current was recorded using a custom MRI-conditional setup, and peak amplitudes were quantified across multiple sequences, including gradient-intensive fMRI protocols. All data were normalized to baseline output and analyzed using Cohen's d to assess the magnitude of MRI-induced current deviation. RESULTS: Monopolar stimulation during MRI exhibited significant current fluctuations, with induced amplitudes ranging from -3.2 to +3.9 mA and frequent polarity inversion. These distortions were sequence dependent and most pronounced during fMRI acquisition. In contrast, bipolar stimulation demonstrated stable output with minimal deviation from programmed parameters. The variability observed in monopolar output was not attributable to impedance shifts and was consistent across both DBS systems tested. CONCLUSIONS: MRI-induced current substantially alters the effective output of monopolar DBS, introducing uncertainty into any concurrent fMRI acquisition. Although functional imaging was not directly performed in human subjects, these findings imply that the observed blood oxygen level-dependent (BOLD) response during monopolar stimulation likely reflects the distorted, not programmed, stimulation. Bipolar configurations avoid this confounder and should be preferred when interpreting fMRI data acquired during DBS.
引用
- Wu C, Slovensky S, Alizadeh M, et al. Induced current effects on functional magnetic resonance imaging interpretability during monopolar deep brain stimulation. J Neurosurg. 2026. PMID: 41687095. doi:10.3171/2025.9.JNS251514
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。