脑室腹腔分流术后早期容积变化与阀门MRI抗性的随机双盲研究
结果发现设4组脑室容积缩小16±9 mL,设8组缩小5±5 mL,组间差异显著;所有MRI均未导致阀门设置改变。
要点
- 脑室容积测定可早期检出分流术后的脑室缩小。
- 阀门阻力设置影响术后早期脑室容积缩小幅度。
- Codman Certas® Plus阀门在临床MRI检查中设置稳定。
核心结论
- 脑室容积测定能检测分流术后早期脑室容积缩小,缩小幅度与分流阀门阻力相关。
- Codman Certas® Plus阀门在临床环境中对MRI诱导的设置改变保持稳定。
研究设计与样本
- 患者接受定量MRI脑室容积测定,分流术前及术后36小时各一次。
- 阀门设置分为4(20例)和8(25例),设置对患者和检查者设盲。
- 3年内共完成156次MRI检查,每次MRI后评估阀门设置。
主要发现
- 设4组术后早期脑室容积缩小16(标准差±9)mL,设8组缩小5(标准差±5)mL,组间差异显著。
- 设4组术后大脑高凸面及旁矢状窦蛛网膜下腔脑脊液间隙狭窄程度显著轻于设8组。
- 所有MRI检查后均未发现阀门设置因MRI而发生改变。
临床怎么用
- 对于特发性正常压力脑积水患者,可采用脑室容积测定早期评估分流效果。
- 阀门阻力设置可能影响术后早期脑室缩小幅度,需结合临床选择。
- 使用Codman Certas® Plus阀门时可减少对MRI导致设置改变的担忧。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Linear radiological measures have low sensitivity to detect changes in ventricular volume in patients with idiopathic normal pressure hydrocephalus. Ventricular volumetry is accurate and sensitive in detecting subtle changes in cerebrospinal fluid volumes. The Codman Certas® Plus is an adjustable shunt valve with 8 settings and resistant to magnetic resonance imaging (MRI)-induced inadvertent adjustments in vitro. The aim of this study was to investigate early volumetric changes in ventricles after ventriculoperitoneal shunting in relation to shunt setting and linear measures. We also wanted to evaluate the MRI-resistance of the Codman Certas® Plus valve in a clinical setting. METHODS: Forty-five idiopathic normal pressure hydrocephalus patients underwent quantitative MRI, including volumetry before and 36 hours after shunting with Codman Certas® Plus valves set to 4 (20 patients) and 8 (25 patients). Valve setting was blinded to patients and examiners and assessed after each MRI. Patients performed in total 156 MRI examinations during 3 years. RESULTS: There was significant difference in change of ventricular volume between groups 4 and 8 early after surgery. Patients with setting 4 had a ventricular volume reduction of 16 (standard deviation ± 9) mL while those with setting 8 had a reduction of 5 (standard deviation ± 5) mL. Constriction of subarachnoid cerebrospinal fluid spaces in cerebral high convexity and parafalcine sulci was significantly less in the setting 4 group postoperatively. There were no MRI-induced changes to valve setting after any MRI. CONCLUSIONS: Ventricular volumetry can detect shunt-induced reduction in ventricle volume early after surgery. The magnitude of reduction is related to shunt valve resistance. The Codman Certas® Plus valve is stable against MRI-induced changes in a clinical setting.
引用
- Holmgren RT, Nilsson M, Georgiopoulos C, et al. A Randomized Double-Blinded Clinical Study of Early Volumetric Changes After Shunt Surgery and MRI-Resistance of the Codman Certas® Plus Shunt Valve. World Neurosurg. 2025. PMID: 40887008. doi:10.1016/j.wneu.2025.124424
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。