文献速递功能神经外科

机械臂与立体定向框架在运动障碍DBS手术中的对比:回顾性队列研究

Robotic arm vs. stereotactic frame in deep brain stimulation surgery for movement disorders: a retrospective cohort study.
PubMed 文献 · Acta Neurochir (Wien) · 2025-08-12

单中心回顾性队列比较机械臂Neuromate与立体定向框架辅助脑深部电刺激(DBS)电极植入。

要点

  • 机器人与框架两组术后3个月UPDRS改善百分比相似(71.4±18 对 72.6%±17)。
  • 围手术期并发症与刺激、药物相关不良反应两组无显著差异。
  • 机器人组电极植入径向误差略小于框架组,解剖影像精度稍优。

核心结论

  • 两种系统初始临床疗效与安全性相似,机器人Neuromate在DBS电极植入的解剖影像精度上略有改善。

主要发现

  • 术后3个月UPDRS改善百分比机器人组71.4±18、框架组72.6%±17,P=0.82。
  • 围手术期并发症机器人组4%、框架组4.3%,P=0.93。
  • 刺激与药物治疗相关不良反应机器人组18%、框架组25%,P=0.53。
  • 径向误差机器人组1.01±0.5 mm、框架组1.32±0.6 mm,P=0.03。

临床怎么用

  • 运动障碍DBS可选机械臂或立体定向框架,两者早期疗效与安全性相当,机器人或带来略优的电极定位精度。

局限性

  • 为单中心回顾性队列研究,样本量有限,非随机分组,结论需前瞻性研究验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Recently, robotic arms have been incorporated into the implantation of electrodes for deep brain stimulation (DBS).This study aimed to determine the accuracy of brain electrode placement, initial clinical efficacy, and safety profile of the robotic arm Neuromate (Renishaw) compared to a stereotactic frame in movement disorders. METHODS: This study involved two retrospective cohorts: one cohort was operated on using a stereotactic frame and the other with a robotic arm. This study was conducted at Barcelona Hospital Clinic. RESULTS: Seventy-seven patients were included, of whom 30 underwent surgery using the robot and 47 using a stereotactic frame. The postoperative improvement percentage of the Unified Parkinson's Disease Rating Scale at 3 months was similar in both groups (robot: 71.4 ± 18 vs. frame: 72.6% ± 17, P = 0.82). There were no significant differences in the perioperative complications (robot: 4% vs. frame: 4.3%, P = 0.93) or in the adverse reactions related to brain stimulation and medical treatment (robot: 18% vs. frame: 25%, P = 0.53). There was a slight improvement in the anatomical-radiological accuracy of brain electrode implantation assisted by the robotic arm, measured using radial error (robot: 1.01 ± 0.5 mm vs. frame: 1.32 ± 0.6 mm, P = 0.03). CONCLUSIONS: Both systems (robotic and stereotactic frame) exhibited similar initial clinical efficacies and safety profiles. The use of the robotic arm Neuromate slightly improved the anatomical-radiological accuracy in the placement of DBS electrodes for movement disorders compared with the stereotactic frame.

来源:Acta Neurochir (Wien),查看原文。版权归原作者所有。

引用

  1. Perera D, Ramos PR, Valldeoriola F, et al. Robotic arm vs. stereotactic frame in deep brain stimulation surgery for movement disorders: a retrospective cohort study. Acta Neurochir (Wien). 2025. PMID: 40794297. doi:10.1007/s00701-025-06618-0