气颅对帕金森病丘脑底核脑深部电刺激术后电极位置的影响
该回顾性分析纳入111例在安徽医科大学第一附属医院接受双侧丘脑底核脑深部电刺激术的帕金森病患者,利用术前MRI与术后CT评估电极位置及气颅体积。研究发现气颅吸收显著影响电极定位,导致电极向前上方移位;脑萎缩程度更高和微电极记录通道数更多与气颅体积增大及电极移位更明显显著相关。
要点
- 气颅吸收显著影响电极位置,导致电极向前上方移位。
- 脑萎缩程度更高和微电极记录通道数更多与气颅体积增大显著相关。
- 减少脑脊液流失和仔细规划微电极记录通道对提高手术精度至关重要。
核心结论
- 气颅在丘脑底核脑深部电刺激术中导致电极移位,减少脑脊液流失和仔细规划微电极记录通道对提高手术精度至关重要。
- 作者指出需要更大样本量和多中心数据的进一步研究来验证这些发现并提高其普遍性。
研究设计与样本
- 回顾性分析纳入111例在安徽医科大学第一附属医院接受双侧丘脑底核脑深部电刺激术的帕金森病患者。
- 利用术前磁共振成像和术后计算机断层扫描评估电极位置和气颅体积。
- 脑影像数据标准化至蒙特利尔神经学研究所空间以进行精确比较。
主要发现
- 气颅吸收显著影响电极定位,导致电极向前和向上移位。
- 更高的脑萎缩程度和更多的微电极记录通道数与气颅体积增加及更明显的电极移位显著相关。
临床怎么用
- 作者认为减少脑脊液流失和仔细规划微电极记录通道对提高手术精度至关重要。
局限性
- 原文指出需要更大样本量和多中心数据的进一步研究来验证这些发现并提高其普遍性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: This study investigates the impact of pneumocephalus on electrode positioning following subthalamic nucleus deep brain stimulation (STN-DBS) in patients with Parkinson disease. METHODS: A retrospective analysis was performed on 111 patients who underwent bilateral STN-DBS at the First Affiliated Hospital of Anhui Medical University. Preoperative magnetic resonance imaging and postoperative computed tomography scans were utilized to assess electrode positions and pneumocephalus volume. The brain imaging data were standardized to the Montreal Neurological Institute space for precise comparison. Statistical analyses were performed to identify factors influencing the volume of pneumocephalus. RESULTS: This study found that pneumocephalus absorption significantly affects electrode positioning, leading to a forward and upward shift. A higher degree of brain atrophy and a higher number of microelectrode recording passages were significantly correlated with increased pneumocephalus volume and more pronounced electrode displacement. CONCLUSIONS: Pneumocephalus plays a critical role in electrode displacement during STN-DBS surgery. Minimizing cerebrospinal fluid loss and careful planning of microelectrode recording passages are essential for improving surgical accuracy. Further studies with larger sample sizes and multicenter data are needed to validate these findings and increase their generalizability.
引用
- Wang Y, Mao H, Kong X, et al. Effects of Pneumocephalus on Electrode Location After Deep Brain Stimulation of the Subthalamic Nucleus in Parkinson Disease. World Neurosurg. 2025. PMID: 40189024. doi:10.1016/j.wneu.2025.123958
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。