无创眼底镜检查作为估算颅内压的工具:大型动物模型
本研究在6头猪中通过有创监测控制颅内压(ICP)从正常(5-15 mmHg)至升高(>20 mmHg),用眼底镜测量视网膜动静脉比(A/V比)。混合效应线性回归显示ICP≥20 mmHg时A/V比与ICP呈显著负相关,ICP<20 mmHg时无变化;ROC分析AUC为0.64(ICP≥20 mmHg)和0.71(ICP>眼内压)。
要点
- ICP≥20 mmHg时A/V比随ICP升高而降低,ICP<20 mmHg时A/V比无变化。
- ICP>眼内压(平均10 mmHg)时A/V比变化与ICP≥20 mmHg时无显著差异。
- AUC为0.64(ICP≥20 mmHg)和0.71(ICP>眼内压),诊断准确性有限。
核心结论
- 结果支持ICP升高与A/V比降低相关的假设。
- ICP>眼内压模型拟合稍好,但因受试者眼内压影响被认为临床相关性低于ICP≥20 mmHg模型。
- 需要整合多因素模型和机器学习的进一步研究来提高眼底镜A/V比的诊断准确性,使其能作为成本有效的筛查工具。
研究设计与样本
- 在6头猪中进行的动物实验,在同一动物体内实现从正常(5-15 mmHg)至升高(>20 mmHg)的受控ICP值。
- 用脑实质内压力监测仪测量ICP,用眼压计测量眼内压(IOP),在基线和预设ICP值时进行眼底镜检查。
主要发现
- 混合效应线性回归显示A/V比与ICP≥20 mmHg呈显著负相关(斜率系数-0.0026734 [95%CI:-0.0039347至-0.0014121],p<0.001)。
- ICP<20 mmHg时A/V比无变化(p=0.987)。
- ICP>IOP(平均IOP 10 mmHg)时结果类似;Wald检验显示ICP>IOP与ICP≥20 mmHg之间无显著差异。
- ROC曲线分析显示ICP≥20 mmHg的AUC为0.64,ICP>IOP的AUC为0.71。
局限性
- ICP>IOP模型因受试者眼内压影响被认为临床相关性低于ICP≥20 mmHg模型。
- 需要进一步研究整合多因素模型和机器学习来提高A/V比经眼底镜的诊断准确性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Intracranial pressure (ICP) monitoring is in most studies considered essential in avoiding secondary brain injury in patients with intracranial pathologies. Invasive monitoring of ICP is accurate but is unavailable in many clinical and prehospital settings. Non-invasive modalities have historically been difficult to implement clinically. The retinal arteriovenous ratio (A/V ratio) has shown promise through its relationship with ICP. This study aimed to further elucidate the relationship between ICP, A/V ratio and the intraocular pressure (IOP) measured with non-invasive fundoscopy in a porcine model. METHODS: We achieved controlled values of ICP ranging from normal (5-15 mmHg) to elevated (> 20 mmHg) within the same animal subject. Six pigs were included. ICP and IOP was measured using an intraparenchymal pressure monitor and a tonometer, respectively. Fundoscopy was performed at baseline and at predefined ICP values. RESULTS: Mixed-effects linear regression revealed a significant inverse correlation between A/V ratio and ICP ≥ 20 mmHg (slope coefficient - 0.0026734 [95%-CI: -0.0039347 - (-0.0014121)], p < 0.001). For ICP < 20 mmHg there was no change in A/V ratio (p = 0.987). Similar results were seen for ICP > IOP with a mean IOP of 10 mmHg. A Wald test showed no significant difference between ICP > IOP and ICP ≥ 20 mmHg. ROC curve analysis revealed an AUC of 0.64 for ICP ≥ 20 mmHg and 0.71 for ICP > IOP. CONCLUSION: The results support the hypothesis that an increase in ICP was associated with a decrease in A/V ratio. Although a slightly better fit, the model of ICP > IOP was deemed less clinically relevant than ICP ≥ 20 mmHg because of the subjects' IOP. Further research integrating multifactorial models and machine learning is needed to enhance the diagnostic accuracy of A/V ratio via fundoscopy, enabling it to serve as a cost-effective screening tool.